Mass Spectrometry in the analysis of Persistent Organic Pollutants Anton Kočan, Jana Klánová Tools for determining chemical structure •IR •UV-VIS •X-ray crystallography •NMR •Mass spectrometry (MS) Kurt Wuthrich a John B. Fenn (1/4) + Koichi Tanaka (1/4) shared the Nobel Price in chemistry for 2002. The latter two for the development of soft desorption ionization methods for mass spectrometric analyses of biological macromolecules Altogether, 5 scientists have been awarded the Nobel Price in connection with mass spectrometry – Joseph Thompson in 1906, Francis Aston in 1922, Wolfgang Paul in 1989, and J. Fenn and K. Tanaka in 2002 A. Kočan, Slovak Medical University Advantages of Mass Spectrometry •Mixtures can be analyzed •A small amount of substance is sufficient Disadvantages of Mass Spectrometry •The evaluation of mass spectra is demanding •It is a destructive method A. Kočan, Slovak Medical University Ion Source Mass Analyzer Ion Detector Vacuum Pumps Inlet Sample Introduction Data System Data Output Mass Spectrometer •All the MS systems compose of the following parts: Ionization, separation of ions and their detection take place in vacuum (~ 10-4 – 10-6 Pa) •Electron multiplier •Photomultiplier •Microchannel plate •Magnetic-sector •Quadrupole •Ion trap •Time-of-flight •Ion-cyclotron resonance •Electron impact (EI) •Chemical ionization (CI) •Electrospray (ESI) •Fast-atom bombardment (FAB) •Laser ionization (LIMS) •Resonance ionization (RIMS) •ˇThermal ionization (TIMS) •Plasma-desorption ionization (PD) •Matrix-assisted laser desorption ionization (MALDI) •GC •LC •Direct inlet A. Kočan, Slovak Medical University Sample Introduction •Direct inlet to the ion source –Introduction of the gaseous sample –Introduction of the liquid or solid sample after their vaporization •Outlet from the chromatographic column –Gas chromatography – packed column (a carrier gas separator is needed) – capillary column (directly to the ion source) –Liquid chromatography (the separation of the mobile phase is essential) HexaCDFs separated by packed column 3% Dexsil 300 on Supelcoport 100/120 2.4 m, 2 mm i.d. capillary column SP-2331 60 m 0.25 mm i.d. 0.2 mm A. Kočan, Slovak Medical University Electron Impact Ionization Source ~70 Volts + _ + _ e- e- e- + + + + + + _ Electron Collector (Trap) Repeller Extraction Plate Filament to Analyzer Inlet Electrons Neutral Molecules Positive Ions The molecules are bombarded by electrons (e-) emitted by a heated filament (W or Re) in vacuum (~ 10-5 – 10-6 mbar) A. Kočan, Slovak Medical University Electron Impact Ionization Source ~70 Volts + _ + _ e- e- e- + + + + + + _ Electron Collector (Trap) Repeller Extraction Plate Filament to Analyzer Inlet Electrons Neutral Molecules Positive Ions The electrons are accelerated towards an anode (electron trap) A. Kočan, Slovak Medical University Electron Impact Ionization Source ~70 Volts + _ + _ e- e- e- + + + + + + _ Electron Collector (Trap) Repeller Extraction Plate Filament to Analyzer Inlet Electrons Neutral Molecules Positive Ions Energy of e- : 15 to 100 eV; 70 eV is usually used for scaning mass spectra A. Kočan, Slovak Medical University Electron Impact Ionization Source ~70 Volts + _ + _ e- e- e- + + + + + + _ Electron Collector (Trap) Repeller Extraction Plate Filament to Analyzer Inlet Electrons Neutral Molecules Positive Ions On their path they collide with gaseous molecules injected into the ion source or eluted from a GC/LC column On average, one ion is produced for 1000 molecules entering the ion source at 70 eV A. Kočan, Slovak Medical University Electron Impact Ionization Source ~70 Volts + _ + _ e- e- e- + + + + + + _ Electron Collector (Trap) Repeller Extraction Plate Filament to Analyzer Inlet Electrons Neutral Molecules Positive Ions 10 to 20 eV out of those 70 eV are transferred to the molecules during the ionization process; Since ~ 10 eV are enough to ionize most organic molecules the excess energy leads to extensive fragmentation; Hence EI is classified as a „hard“ ionization technique The fragmentation gives structural information A. Kočan, Slovak Medical University Electron Ionization •One of the oldest and simplest methods (1930 – 1950) •Applicable for vaporizable samples only, e.g. compounds with low molecular weight (< 1000 Da), less polar organic molecules •Advantages: –Unimolecular fragmentation (ions spend 1 ms in EI source) –Reproducible technique •Disadvantages: –Sometimes M+. is not observed due to fragmentation –Almost impossible distinguish between isomers (GC separation needed) –Some compounds can undergo thermal decomposition before ionization because of high temperatures used to vaporize the sample –Inappropriate to too involatile compounds A. Kočan, Slovak Medical University Schematic of Electron Ionization Ion-sourceMAT Ionization chamber A. Kočan, Slovak Medical University Positive Chemical Ionization •Developed in the 1960s. •Gaseous ions such as CH5+, C2H5+, C4H9+, NH4+, H3+ generated by electron impact from a large excess of a reagent gas, such as CH4, NH3, H2, or i-C4H10, interact with neutral molecules that may ionize. •Generally, the amount of fragments is much less than in EI since little internal energy is imparted on the ionized molecule; Thus, the important molecular ion can be determined. •Hence, CI is termed as „soft“ ionization technique. A. Kočan, Slovak Medical University Electrospray MALDI ion trajectory not in register (too heavy) Ion Source Detector ion trajectory not in register (too light) ion trajectory in register S N Electromagnet Magnetic Sector Mass Analyzer HPIM3647 A. Kočan, Slovak Medical University High Resolutionn MS Autospec-Ultima •3-sector instrument (2 electrostatic sectors, 1 magnetic sector) VGAutospecMdiagram •Mattauch-Herzog geometry VGAutospecMphoto1 A. Kočan, Slovak Medical University P1220003 Dual HRGC / HRMS System DB5-ms Rtx-2330 A. Kočan, Slovak Medical University Detector Ion Source + _ _ DC and AC Voltages resonant ion non-resonant ion + Quadrupole Mass Filter quad1 •Consists of 4 parallel metal rods. •Two opposite rods have an applied potential combined from DC and AC voltages. •A mass spectrum is obtained by monitoring the ions passing through the quadrupole filter as the voltages or frequency on the rods are varied. •For given dc and ac voltages, only ions of a certain mass-to-charge ratio pass through the filter and all other ions are thrown out of their original path. •The voltages affect the trajectory of ions traveling down the flight path centered between the rods. A. Kočan, Slovak Medical University TOF MS schematic Time-Of-Flight Mass Spectrometry (TOFMS) •It uses differences in transit time through a drift region to separate ions of different masses •It operates in a pulsed mode so ions must be produced or extracted in pulses •An electric field accelerates all ions into a field-free drift region with the same initial kinetic energy for all the ions produced •Since the ion kinetic energy is 0.5mv2, lighter ions have a higher velocity than heavier ions and reach the detector sooner (e.g., ions of m/z 500 arrive in ~ 15 ms and m/z 50 in ~ 4.6 ms Schematic of gas analysis system Photo of TOF-MS for gases •By TOF-MS, up to 50 000 full spectra can be measured in a second •Since full spectra are available, peak deconvolution software enabling to differentiate non-separated GC peaks may be applied •The TOF ultra-fast scanning is suitable for fast GC where peak widths can be much less then a second A. Kočan, Slovak Medical University Ion Trap Mass Spectrometry •The advantages of the ion-trap mass spectrometer include compact size, and the ability to trap and accumulate ions to increase the signal-to-noise ratio of a measurement. picture of w. paul Wolfgang Paul 1989 Nobel Price for Physics „for the development of the ion trap technique“ •This technique can be used easily in the MS/MS (MSn) mode •The ion-trap analyzer consists of 3 electrodes with hyperbolic surfaces to trap ions in a small volume – the central ring electrode and 2 adjacent endcap electrodes. A mass spectrum is obtained by changing the electrode voltages to eject the ions from the trap. A. Kočan, Slovak Medical University To Amplifier -3 kV 0 V Dynode Electron Multiplier Continuous-dynode electron multiplier Discrete-dynode electron multiplier typical current gain is 106 A. Kočan, Slovak Medical University fil-hp §Mean free path of molecules Why needed? Vacuum System üRough pumps diffusion-180 üDiffusion pumps üTurbomolecular pumps §Ion-molecular reactions §Interferences in mass spectra §Glowing/sparking in the high-voltage area §Contamination of the ion source §Burning of the filament HPIM3646 HPIM3650 A. Kočan, Slovak Medical University Mass Spectrum m/z (mass-to-charge ratio) Relative abundance (%) 100 Molecular ion M+. Base peak [M-X]+. Fragment ion [M-Y]+. There are now Internet-accessible databases containing mass spectra of thousands of compounds, e.g.: http://www.aist.go.jp/RIODB/SDBS/menu-e.html http://webbook.nist.gov A. Kočan, Slovak Medical University Segment of the Real Signal (not computer-processed) of a Mass Spectrum (n-cethyl amine – C16H33NH2) m/z •Such a Gauss-shaped peaks are processed by a computer program to be converted to the line spectrum 142 156 170 184 198 212 226 241 m/z A. Kočan, Slovak Medical University GC-MS Uses •Identification and quantification of volatile and semivolatile organic compounds in complex mixtures •Determination of molecular weights and (sometimes) elemental composition of unknown organic compounds in complex mixtures •Structural determination of unknown organic compounds in complex mixtures both by matching their mass spectra with reference spectra and by a priori spectral interpretation When to Use HRMS ? •For the determination of the exact masses of ions in spectra (peak matching) – suitable for the determination of elemental composition •Decreasing interferences caused by the ions of co-extracted compounds or by GC column bleeding and thus, to improve considerably the signal-to-noise ratio A. Kočan, Slovak Medical University Fragmentacia_PCDFs Fragmentacia_PCDDs Fragmentation schemes under EI conditions Polychlorinated dibenzo-p-dioxins Polychlorinated dibenzofurans * fragmentation for lower chlorinated congeners A. Kočan, Slovak Medical University Demonstration of Mass Spectrometric Resolution (10% Valley Definition) R10% = m / Dm A. Kočan, Slovak Medical University •In general, more ions have the same nominal mass •To distinguish between them certain MS resolution is needed •For example, to separate these 2 ions we need a resolution of 5 124 R = (122.060585 – 122.036776) / 122 = 5 124 A. Kočan, Slovak Medical University untitled Relative Abundances of the Combinations of Halogen Isotopes •Fluorine and iodine are monoisotopic •However, chlorine is present in the nature as 35Cl (75.8%) and 37Cl (24.2%), and bromine as 79Br (50.5%) and 81Br (49.5%) •Therefore molecular ions (or fragment ions) containing various numbers of Cl and/or Br atoms give rise to typical patterns spaced 2 amu apart based on a combination of (a+b)n; for Cl a is 75.8%, b 24.2% and n is the number of chlorine substituents in a molecule A. Kočan, Slovak Medical University M+. [M+2]+. [M+4]+. [M+6]+. 320 [M-Cl]+. [M-COCl]+. Upper Segment of 2378-TCDD Mass Spectrum A cluster created by four Cl-substituents 4×35Cl 3×35Cl 1×37Cl 2×35Cl 2×37Cl 1×35Cl 3×37Cl 4×37Cl A. Kočan, Slovak Medical University What is the SCAN Mode in Mass Spectrometry ? •The scanning mode provides mass spectra. They are recorded (scanned) at regular intervals (typically 0.5 – 1 /s; much faster if TOFMS is used) during the GC separation and stored in the instrument data system for subsequent qualitative or quantitative evaluation. •From mass spectra, it is often possible to deduce structural features (mass spectral interpretation) but this requires experience and can be very time-consuming, particularly as a complex mixture might contain hundreds of components. •The spectra can also be compared with those stored in mass spectral libraries. Although library searching is a very useful and timesaving technique, it is important to remember that such searches do not identify compounds – analysts do! A. Kočan, Slovak Medical University What is the SIM (or MID) Mode in Mass Spectrometry ? •SIM (Selected Ion Monitoring) or MID (Multiple Ion Detection) is much more sensitive technique suitable for trace quantitative analysis. Here, instead of scanning a whole spectrum, only a few ions (generally, the most abundant but characteristic selected from the mass spectrum) are detected during the GC run. •This can result in as much as a 500-fold increase in sensitivity, at the expense of selectivity. Depending on the analyte, low picogram to even low femtogram amounts can be measured using this powerful technique. •Stable isotope-labeled internal standards can be employed. f3 HRMS/LRMS-SIM chromatogram from the analysis of 2378-TCDD in a soil extract by the isotope dilution method 2,3,7,8-TCDD 13C12-2,3,7,8-TCDD M+. A=1392 (82%) m/z 319.9 m/z 321.9 m/z 323.9 [M+2]+. A=1688 (100 %) [M+4]+. A=829 (49%) M+. A=1936 (80%) m/z 331.9 m/z 333.9 m/z 335.9 [M+2]+. A=2415 (100%) [M+4]+. A=1198 (50%) A. Kočan, Slovak Medical University The Use of the Isotope Dilution HRGC/HRMS in the SIM Mode 13C-labeled standards 13C-labeled standards M+. M+. [M+2]+. [M+2]+. TetraCDDs A. Kočan, Slovak Medical University PeCDF_flyash A. Kočan, Slovak Medical University Dioxins and Related Compounds Analysis (1) Types of samples: §Emissions from combustion processes (stack aerosol, flyash) §Wastes (obsolete chemicals, disposal sites) §Environmental samples (air, soil, water sediment, biota) §Foods (meat, fish, milk, eggs, oils) §Human biological material (milk, blood, adipose tissue) A. Kočan, Slovak Medical University Sampling train Rinsing of sampling device Extraction Concentration Silica Silica + H2SO4 Silica Silica + NaOH Silica Silica + AgNO3 Sampling standards Extraction standards Basic Alumina Super I column Active carbon column Concentration GC/MS Syringe standards Flow chart of a clean-up procedure for stack emission samples A. Kočan, Slovak Medical University Stack1 Dioxins and Related Compounds Analysis (2) Sampling: §The collection of samples should represent the situation, process or the species studied §Attention must be paid to the appropriate timing of collections of samples, their frequency and site layout §Sampling media, particularly those for air and emission sampling should be spiked before sampling with selected isotope labeled surrogates (13C-sampling standards) to make corrections of losses of analytes during the sampling A. Kočan, Slovak Medical University Dioxins and Related Compounds Analysis (3) Sampling extraction & cleanup: §Addition of isotope labeled standards (13C) before extraction §Solvent extraction (Soxhlet, ASE, ultrasonic, SPE, SFE, MASE) §Removal of matrix substances (lipids, oils, etc.) – H2SO4, GPC, SPM dialysis §Removal of interfering coextracted compounds (PCBs, PCDPEs, etc.) – acid/base modified silica, alumina, active carbon, semi-preparative HPLC) §Concentration of the analytes SPE of blood serum Cleanup on silica/H2SO4 columns A. Kočan, Slovak Medical University Schematic of a semi-automatic equipment for sample cleanup Semi-automated clean-up equipment A. Kočan, Slovak Medical University Dioxins and Related Compounds Analysis (4) GC separation: §Non-polar stationary phase (e.g. DB-5) – used for the samples of animal origin and higher chlorinated congeners §Polar phase (e.g. SP-2330) – used for environmental samples (good separation but shorter lifetime) §Splitless, on-column or large-volume injection §Direct connection of the column to the ion source out of splitless injector to mass spectrometer A. Kočan, Slovak Medical University HxCDF_flyash Identification Criteria (according to the EN 1948 Standard) §The isotope ratio between the ions monitored shall match the theoretical value within ± 20 % §The retention time of a native 2,3,7,8-isomer shall be within a time window of (0s;+3s) based on the retention time of the corresponding 13C12-labelled isomer in the sample. For hepta- and octachloro-congeners deviations of (-2s;+3s) are acceptable §The GC column shall separate 2,3,7,8-congeners from interfering congeners with a 90 % valley relative to the highest peak. 2,3,7,8-TCDF shall be separated from all other interfering isomers within a 25 % valley Dioxins and Related Compounds Analysis (5) M+2 M+4 A. Kočan, Slovak Medical University Group Ion % Abundance Natives 13C12-labelled TetraCDFs TetraCDDs [M]+. [M+2]+. [M]+. [M+2]+. 77.55 100 77.55 100 303.9016 305.8987 319.8965 321.8935 315.9418 317.9390 331.9398 333.9338 PentaCDFs PentaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 64.15 100 64.15 339.8597 341.8567 355.8546 357.8516 351.9000 353.8970 367.8948 369.8918 HexaCDFs HexaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 80.54 100 80.54 373.8208 375.8178 389.8156 391.8126 385.8610 387.8581 401.8558 403.8528 HeptaCDFs HeptaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 96.52 100 96.52 407.7818 409.7789 423.7766 425.7737 419.8220 421.8192 435.8169 437.8140 OctaCDF OctaCDD [M+2]+. [M+4]+. [M+2]+. [M+4]+. 88.89 100 88.89 100 441.7428 443.7398 457.7377 459.7348 453.7830 455.7801 469.7779 471.7750 Mass of ions monitored for tetra- to octaCDDs/Fs when HRMS is applied Group Ion % Abundance Natives 13C12-labelled TetraCDFs TetraCDDs [M]+. [M+2]+. [M]+. [M+2]+. 77.55 100 77.55 100 303.9016 305.8987 319.8965 321.8935 315.9418 317.9390 331.9398 333.9338 PentaCDFs PentaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 64.15 100 64.15 339.8597 341.8567 355.8546 357.8516 351.9000 353.8970 367.8948 369.8918 HexaCDFs HexaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 80.54 100 80.54 373.8208 375.8178 389.8156 391.8126 385.8610 387.8581 401.8558 403.8528 HeptaCDFs HeptaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 96.52 100 96.52 407.7818 409.7789 423.7766 425.7737 419.8220 421.8192 435.8169 437.8140 OctaCDF OctaCDD [M+2]+. [M+4]+. [M+2]+. [M+4]+. 88.89 100 88.89 100 441.7428 443.7398 457.7377 459.7348 453.7830 455.7801 469.7779 471.7750 Mass of ions monitored for tetra- to octaCDDs/Fs when HRMS is applied Group Ion % Abundance Natives 13C12-labelled TetraCDFs TetraCDDs [M]+. [M+2]+. [M]+. [M+2]+. 77.55 100 77.55 100 303.9016 305.8987 319.8965 321.8935 315.9418 317.9390 331.9398 333.9338 PentaCDFs PentaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 64.15 100 64.15 339.8597 341.8567 355.8546 357.8516 351.9000 353.8970 367.8948 369.8918 HexaCDFs HexaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 80.54 100 80.54 373.8208 375.8178 389.8156 391.8126 385.8610 387.8581 401.8558 403.8528 HeptaCDFs HeptaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 96.52 100 96.52 407.7818 409.7789 423.7766 425.7737 419.8220 421.8192 435.8169 437.8140 OctaCDF OctaCDD [M+2]+. [M+4]+. [M+2]+. [M+4]+. 88.89 100 88.89 100 441.7428 443.7398 457.7377 459.7348 453.7830 455.7801 469.7779 471.7750 Mass of ions monitored for tetra- to octaCDDs/Fs when HRMS is applied Group Ion % Abundance Natives 13C12-labelled TetraCDFs TetraCDDs [M]+. [M+2]+. [M]+. [M+2]+. 77.55 100 77.55 100 303.9016 305.8987 319.8965 321.8935 315.9418 317.9390 331.9398 333.9338 PentaCDFs PentaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 64.15 100 64.15 339.8597 341.8567 355.8546 357.8516 351.9000 353.8970 367.8948 369.8918 HexaCDFs HexaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 80.54 100 80.54 373.8208 375.8178 389.8156 391.8126 385.8610 387.8581 401.8558 403.8528 HeptaCDFs HeptaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 96.52 100 96.52 407.7818 409.7789 423.7766 425.7737 419.8220 421.8192 435.8169 437.8140 OctaCDF OctaCDD [M+2]+. [M+4]+. [M+2]+. [M+4]+. 88.89 100 88.89 100 441.7428 443.7398 457.7377 459.7348 453.7830 455.7801 469.7779 471.7750 Mass of ions monitored for tetra- to octaCDDs/Fs when HRMS is applied Group Ion % Abundance Natives 13C12-labelled TetraCDFs TetraCDDs [M]+. [M+2]+. [M]+. [M+2]+. 77.55 100 77.55 100 303.9016 305.8987 319.8965 321.8935 315.9418 317.9390 331.9398 333.9338 PentaCDFs PentaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 64.15 100 64.15 339.8597 341.8567 355.8546 357.8516 351.9000 353.8970 367.8948 369.8918 HexaCDFs HexaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 80.54 100 80.54 373.8208 375.8178 389.8156 391.8126 385.8610 387.8581 401.8558 403.8528 HeptaCDFs HeptaCDDs [M+2]+. [M+4]+. [M+2]+. [M+4]+. 100 96.52 100 96.52 407.7818 409.7789 423.7766 425.7737 419.8220 421.8192 435.8169 437.8140 OctaCDF OctaCDD [M+2]+. [M+4]+. [M+2]+. [M+4]+. 88.89 100 88.89 100 441.7428 443.7398 457.7377 459.7348 453.7830 455.7801 469.7779 471.7750 Mass of ions monitored for tetra- to octaCDDs/Fs when HRMS is applied Quantification criteria (according to the EN 1948 Standard) §The S/N ratio of the raw data shall be at least 3:1 for the signal used for quantification §The calibration is carried out with at least 5 calibration solutions §The measuring range shall be linear §Daily calibration checks shall be run §HRMS at 6000 to 10000 resolution can be used if the absence of interferences is documented §Other techniques can be used if show meeting the criteria Dioxins and Related Compounds Analysis (6) A. Kočan, Slovak Medical University •This conversion is based on the assumption that all the 2,3,7,8-substituted PCDDs and PCDFs (17 cong.), as well as the dioxin-like PCBs (12 cong.), bind to the same receptor, the Ah receptor, and show comparable qualitative (toxic) effects, but with different potencies TEQ = (PCDDi × TEFi) + (PCDFi × TEFi ) + (PCBi × TEFi ) Conversion of Analytical Results into the Toxic Equivalent (TEQ) •These differences in toxicity are expressed in the toxic equivalency factors (TEFs) •TEF of the most toxic 2378-TCDD = 1 Congener I-TEF WHO-TEF Congener I-TEF WHO-TEF 2378-TCDD 1 1 2378-TCDF 0.1 0.1 12378-PeCDD 0.5 1 23478-PeCDF 0.5 0.5 123478-HxCDD 0.1 0.1 12378-PeCDF 0.05 0.05 123678-HxCDD 0.1 0.1 123478-HxCDF 0.1 0.1 123789-HxCDD 0.1 0.1 123789-HxCDF 0.1 0.1 1234678-HpCDD 0.01 0.01 123678-HxCDF 0.1 0.1 OCDD 0.001 0.0001 234678-HxCDF 0.1 0.1 1234678-HpCDF 0.01 0.01 1234789-HpCDF 0.01 0.01 OCDF 0.001 0.0001 A. Kočan, Slovak Medical University Relative Response Factors Quantification Ci = Qi,12C / Vnr CT = S (Ci × I-TEFi ) Extraction Standards Recovery Sampling Standards Recovery Calculations Regarding Dioxin Determination by the Isotope Dilution Method A. Kočan, Slovak Medical University