NOVOTNÝ, Karel, Lucie KRAJCAROVÁ, Aleš HRDLIČKA, Viktor KANICKÝ, Jozef KAISER, Radomír MALINA and David HAHN. Ablation Chamber for Simultaneous DP-LIBS and Laser Ablation ICP-OES/MS. In SCIX 2012. 2012.
Other formats:   BibTeX LaTeX RIS
Basic information
Original name Ablation Chamber for Simultaneous DP-LIBS and Laser Ablation ICP-OES/MS
Name in Czech Ablační komora pro simultánní DP-LIBS a Laserovou ablaci ve spojení s ICP-OES/MS
Authors NOVOTNÝ, Karel (203 Czech Republic, guarantor, belonging to the institution), Lucie KRAJCAROVÁ (203 Czech Republic), Aleš HRDLIČKA (203 Czech Republic, belonging to the institution), Viktor KANICKÝ (203 Czech Republic, belonging to the institution), Jozef KAISER (203 Czech Republic), Radomír MALINA (203 Czech Republic) and David HAHN (840 United States of America).
Edition SCIX 2012, 2012.
Other information
Original language English
Type of outcome Conference abstract
Field of Study 10406 Analytical chemistry
Country of publisher United States of America
Confidentiality degree is not subject to a state or trade secret
RIV identification code RIV/00216224:14740/12:00061408
Organization unit Central European Institute of Technology
Keywords (in Czech) DP-LIBS; Laserová ablace; ICP-OES/MS; ablační komora
Keywords in English DP-LIBS; Laser ablation; ICP-OES/MS; ablation chamber
Tags International impact
Changed by Changed by: doc. Mgr. Karel Novotný, Ph.D., učo 17748. Changed: 29/3/2013 08:41.
Abstract
A special interaction chamber was designed to be compatible with the New Wave UP MACRO Laser Ablation System and to fulfill all demands for simultaneous measurements double pulse laser-induced breakdown spectroscopy (DP-LIBS) and laser ablation coupled to ICP optical emission or mass spectrometry. For single pulse LIBS only two orthogonal fused silica windows are used, the first from the top for the ablation laser beam and the second one from the front for collecting the light of plasma. For DP-LIBS the third window from the side is needed being orthogonal to the previous two. For Laser ablation ICP OES/MS a flow of transporting gas has to be implemented. The inner space of the chamber was optimized for low volume and quick flush of gas, which is crucial for the method. The gas enters the chamber through three oblique jets on the top and leaves to the ICP on the back side. The chamber is equipped with a vertically adjustable sample holder to adapt various sample heights and exchangeable inner rings allowing to various sample diameters to be altered and to minimize the inner volume.
Abstract (in Czech)
Speciální ablační komora byla navržena tak, aby splňovala veškeré požadavky pro provedení simultánního měření technikou dvoupulzní spektroskopie laserem buzeného plazmatu DP-LIBS a laserové ablace ve spojení s optickou nebo hmotností spektrometrií indukčně vázaného plazmatu a současně byla plně kompatibilní s komerčním ablačním systémem firmy New Wave - UP 266 MACRO. Pro techniku jednopulzní LIBS spektrometrie jsou využívána pouze dvě na sebe kolmé okénka komory: první pro vstup ablačního laseru a druhé pro sběr záření laserem buzeného mikroplazmatu. Pro dvoupulzní techniku je zapotřebí třetí okénko pro vstup dalšího laserového pulzu, používaného pro reexcitaci laserem buzeného mikroplazmatu. Pro techniku LA-ICP-OES/MS je komora proplachována nosným plynem. Vnitřní objem komory byl optimalizován s ohledem na co nejrychlejší a nejefektivnější proplachování komory. Komora je dále vybavena posuvným stolkem pro přesné nastavení polohy vzorku.
Links
ED1.1.00/02.0068, research and development projectName: CEITEC - central european institute of technology
ME08002, research and development projectName: Použití spektroskopie laserem buzeného plazmatu (LIBS) a laserové ablace ve spojení s ICP-OES/MS spektroskopií pro chemické mapování biominerálů
Investor: Ministry of Education, Youth and Sports of the CR, Utilization of the laser induced breakdown spectroscopy (LIBS) and laser ablation ICP-OES/MS spectroscopy for chemical mapping of biominerals, Research and Development Programme KONTAKT (ME)
PrintDisplayed: 20/7/2024 03:36