KRATOCHVÍL, Tomáš, Magdalena ZVOLSKÁ, Miloslav POUZAR, Karel NOVOTNÝ and Tomáš ČERNOHORSKÝ. The use of Laser-Induced Breakdown Spectroscopy for the Quantitative Analysis of Fluorine. In ESAS 2012. 2012.
Other formats:   BibTeX LaTeX RIS
Basic information
Original name The use of Laser-Induced Breakdown Spectroscopy for the Quantitative Analysis of Fluorine
Name in Czech Použití spektroskopie laserem buzeného plazmatu pro kvantitativní analýzu fluoru
Authors KRATOCHVÍL, Tomáš (203 Czech Republic), Magdalena ZVOLSKÁ (203 Czech Republic), Miloslav POUZAR (203 Czech Republic), Karel NOVOTNÝ (203 Czech Republic, guarantor, belonging to the institution) and Tomáš ČERNOHORSKÝ (203 Czech Republic).
Edition ESAS 2012, 2012.
Other information
Original language English
Type of outcome Conference abstract
Field of Study 10406 Analytical chemistry
Country of publisher Slovakia
Confidentiality degree is not subject to a state or trade secret
RIV identification code RIV/00216224:14310/12:00061415
Organization unit Faculty of Science
Keywords (in Czech) LIBS; kvantitativní analýza; fluor
Keywords in English LIBS; Quantitative analysis; Fluorine
Tags International impact
Changed by Changed by: doc. Mgr. Karel Novotný, Ph.D., učo 17748. Changed: 22/10/2012 18:45.
Abstract
The aim of this work was to develop an effective, rapid and simple method for the quantitative analysis of fluorine in Glass Ionomer Cement (GIC). Calibration curves describing the relationship between the fluorine concentration and the corresponding intensity of LIBS signal, were constructed for two LIBS devices, with and without He flow also. F(I) spectral line of 685.6 nm was found to be considerably sensitive and provided the linear calibration. The best LOD value was obtained for the device with the short gate width (5 microseconds) in He atmosphere.
Abstract (in Czech)
Úkolem práce bylo vyvinou efektivní, rychlou a jednoduchou metodu pro kvantitativní stanovení fluoru ve skleněném ionomerním cementu (GIC). Kalibrační křivky vyjadřující vztah mezi koncentrací fluoru a odpovídající intenzitou LIBS signálu byly vytvořeny pro dva různé typy instrumentálního uspořádání a to jak v atmosféře helia tak bez ní. Lineární odezva s dostatečnou citlivostí byla pozorována pro atomovou emisní čáru fluoru F(I) 685,6 nm. Nejlepší hodnoty detekčních limitů bylo dosaženo se zařízením s krátkou integrační dobou (5 mikrosekund) v prostředí heliové atmosféry.