FOLTYNOVÁ, Pavla, Antonín BEDNAŘÍK, Viktor KANICKÝ and Jan PREISLER. Diode laser thermal vaporization ICP MS with a simple tubular cell for determination of lead and cadmium in whole blood. Journal of Analytical Atomic Spectrometry. England: Royal Society of Chemistry, 2014, vol. 29, No 9, p. 1585-1590. ISSN 0267-9477. Available from: https://dx.doi.org/10.1039/C4JA00113C.
Other formats:   BibTeX LaTeX RIS
Basic information
Original name Diode laser thermal vaporization ICP MS with a simple tubular cell for determination of lead and cadmium in whole blood
Authors FOLTYNOVÁ, Pavla (203 Czech Republic, belonging to the institution), Antonín BEDNAŘÍK (203 Czech Republic, belonging to the institution), Viktor KANICKÝ (203 Czech Republic, belonging to the institution) and Jan PREISLER (203 Czech Republic, guarantor, belonging to the institution).
Edition Journal of Analytical Atomic Spectrometry, England, Royal Society of Chemistry, 2014, 0267-9477.
Other information
Original language English
Type of outcome Article in a journal
Field of Study 10406 Analytical chemistry
Country of publisher United Kingdom of Great Britain and Northern Ireland
Confidentiality degree is not subject to a state or trade secret
WWW URL
Impact factor Impact factor: 3.466
RIV identification code RIV/00216224:14740/14:00073828
Organization unit Central European Institute of Technology
Doi http://dx.doi.org/10.1039/C4JA00113C
UT WoS 000341064300004
Keywords in English Diode laser thermal vaporization; DLTV; ICP MS; dried blood spot; lead; cadmium
Tags kontrola MP, podil, rivok
Tags International impact, Reviewed
Changed by Changed by: prof. Mgr. Jan Preisler, Ph.D., učo 45329. Changed: 15/2/2018 22:12.
Abstract
A new low-cost device suitable for the recently presented sample introduction technique to inductively coupled plasma mass spectrometry (ICP MS), diode laser thermal vaporization (DLTV), is described. The device performance was optimized, compared to a commercial ablation system and applied to the determination of lead and cadmium in whole human blood without any sample treatment. A prearranged calibration set was printed on the paper strip using a piezoelectric dispenser.