HEJČ, Jakub, Martin VÍTEK, Marina RONZHINA, Marie NOVÁKOVÁ a Jana KOLÁŘOVÁ. A Wavelet-Based ECG Delineation Method: Adaptation to an Experimental Electrograms with Manifested Global Ischemia. Cardiovascular Engineering and Technology. New York: Springer Publishing Company, roč. 6, č. 3, s. 364-375. ISSN 1869-408X. doi:10.1007/s13239-015-0224-z. 2015.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název A Wavelet-Based ECG Delineation Method: Adaptation to an Experimental Electrograms with Manifested Global Ischemia
Autoři HEJČ, Jakub (203 Česká republika), Martin VÍTEK (203 Česká republika), Marina RONZHINA (203 Česká republika), Marie NOVÁKOVÁ (203 Česká republika, garant, domácí) a Jana KOLÁŘOVÁ (203 Česká republika).
Vydání Cardiovascular Engineering and Technology, New York, Springer Publishing Company, 2015, 1869-408X.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 30105 Physiology
Stát vydavatele Spojené státy
Utajení není předmětem státního či obchodního tajemství
Kód RIV RIV/00216224:14110/15:00081012
Organizační jednotka Lékařská fakulta
Doi http://dx.doi.org/10.1007/s13239-015-0224-z
UT WoS 000380356800014
Klíčová slova anglicky ECG delineation; Electrogram; Ischemia; Isolated rabbit heart; Wave detection; Wavelet transform
Štítky EL OK
Příznaky Recenzováno
Změnil Změnila: Ing. Mgr. Věra Pospíšilíková, učo 9005. Změněno: 10. 9. 2015 10:20.
Anotace
We present a novel wavelet-based ECG delineation method with robust classification of P wave and T wave. The work is aimed on an adaptation of the method to long-term experimental electrograms (EGs) measured on isolated rabbit heart and to evaluate the effect of global ischemia in experimental EGs on delineation performance. The algorithm was tested on a set of 263 rabbit EGs with established reference points and on human signals using standard Common Standards for Quantitative Electrocardiography Standard Database (CSEDB). On CSEDB, standard deviation (SD) of measured errors satisfies given criterions in each point and the results are comparable to other published works. In rabbit signals, our QRS detector reached sensitivity of 99.87% and positive predictivity of 99.89% despite an overlay of spectral components of QRS complex, P wave and power line noise. The algorithm shows great performance in suppressing J-point elevation and reached low overall error in both, QRS onset (SD = 2.8 ms) and QRS offset (SD = 4.3 ms) delineation. T wave offset is detected with acceptable error (SD = 12.9 ms) and sensitivity nearly 99%. Variance of the errors during global ischemia remains relatively stable, however more failures in detection of T wave and P wave occur. Due to differences in spectral and timing characteristics parameters of rabbit based algorithm have to be highly adaptable and set more precisely than in human ECG signals to reach acceptable performance.
Návaznosti
GAP102/12/2034, projekt VaVNázev: Analýza vztahu mezi elektrickými ději a průtokem krve u srdečních komor
Investor: Grantová agentura ČR, Analýza vztahu mezi elektrickými ději a průtokem krve u srdečních komor
MUNI/A/0957/2013, interní kód MUNázev: Studium kardiovaskulárních regulací ve zdraví i nemoci (Akronym: SKAREZNEM)
Investor: Masarykova univerzita, Studium kardiovaskulárních regulací ve zdraví i nemoci, DO R. 2020_Kategorie A - Specifický výzkum - Studentské výzkumné projekty
VytisknoutZobrazeno: 20. 4. 2024 03:58