Detailed Information on Publication Record
2017
A Thermographic Comparison of Irreversible Electroporation and Radiofrequency Ablation
BERNARD, Vladan, Tomáš ANDRAŠINA, D. ČERVINKA, J. MARTIŠ, P. PROCHÁZKA et. al.Basic information
Original name
A Thermographic Comparison of Irreversible Electroporation and Radiofrequency Ablation
Name in Czech
Termografické srovnání radiofrekvenční ablace a irreverzibilní elektroporace
Authors
BERNARD, Vladan (203 Czech Republic, guarantor, belonging to the institution), Tomáš ANDRAŠINA (703 Slovakia), D. ČERVINKA (203 Czech Republic), J. MARTIŠ (203 Czech Republic), P. PROCHÁZKA (203 Czech Republic), Vojtěch MORNSTEIN (203 Czech Republic, belonging to the institution) and Vlastimil VÁLEK (203 Czech Republic)
Edition
IRBM, New York, Elsevier Science INC. 2017, 1959-0318
Other information
Language
English
Type of outcome
Článek v odborném periodiku
Field of Study
20601 Medical engineering
Country of publisher
United States of America
Confidentiality degree
není předmětem státního či obchodního tajemství
Impact factor
Impact factor: 0.897
RIV identification code
RIV/00216224:14110/17:00095628
Organization unit
Faculty of Medicine
UT WoS
000398761500004
Keywords in English
Balloon catheter; Heating effect; Infrared camera; Irreversible electroporation; Radiofrequency ablation; Temperature
Tags
Tags
International impact, Reviewed
Změněno: 15/3/2018 17:20, Soňa Böhmová
Abstract
V originále
Radiofrequency ablation and irreversible electroporation are both interventional procedures used in treating cancer diseases generally and may be used in treatment of liver cancer. The two differ in terms of their action and thermal effects. The present study examined thermal effects of radiofrequency ablation and modified irreversible electroporation ex vivo in porcine liver. Standard radiofrequency ablation was used during the experiments, as was a unique, experimental high-voltage generator of the research team's own construction coupled with a newly designed balloon electrode catheter for irreversible electroporation. The tested balloon catheter was newly designed for application in the biliary tract. The temperature of porcine liver tissue was monitored without contact by infrared camera. The results show significantly different temperature of tissue around electrodes for radiofrequency ablation and irreversible electroporation. The data obtained from radiofrequency ablation with maximum tested output power of 10 watts show a symmetrical area of tissue near active electrodes with temperature increased by more than 30 C. Irreversible electroporation methods showed temperature increasing approximately by 5 C. To summarize, the thermal research showed that the difference between standard RFA and innovated IRE with the balloon catheter lies not just in the differing mechanism of action, as is well known, but also in its thermal effects. There was documented that tissue temperature was much higher for RFA than for IRE, particularly for the maximum tested output parameters. The applicability of the newly designed balloon catheters was confirmed.
Links
NV15-32484A, research and development project |
|