ŠEVČÍKOVÁ, Jana, Denisa BÁRTKOVÁ, Monika PÁVKOVÁ GOLDBERGOVÁ, Monika KUBĚNOVÁ, JIří ČERMÁK, Jan FRENZEL, Adam WEISER a Antonín DLOUHÝ. On the Ni-Ion release rate from surfaces of binary NiTi shape memory alloys. Online. Applied Surface Science. Amsterdam: Elsevier Science BV., 2018, roč. 427, 1 January 2018, s. 434-443. ISSN 0169-4332. Dostupné z: https://dx.doi.org/10.1016/j.apsusc.2017.08.235. [citováno 2024-04-23]
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název On the Ni-Ion release rate from surfaces of binary NiTi shape memory alloys
Autoři ŠEVČÍKOVÁ, Jana (203 Česká republika, domácí), Denisa BÁRTKOVÁ (203 Česká republika), Monika PÁVKOVÁ GOLDBERGOVÁ (203 Česká republika, garant, domácí), Monika KUBĚNOVÁ (203 Česká republika), JIří ČERMÁK (203 Česká republika), Jan FRENZEL (203 Česká republika), Adam WEISER (203 Česká republika) a Antonín DLOUHÝ (203 Česká republika)
Vydání Applied Surface Science, Amsterdam, Elsevier Science BV. 2018, 0169-4332.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 30105 Physiology
Stát vydavatele Nizozemské království
Utajení není předmětem státního či obchodního tajemství
Impakt faktor Impact factor: 5.155
Kód RIV RIV/00216224:14110/18:00100744
Organizační jednotka Lékařská fakulta
Doi http://dx.doi.org/10.1016/j.apsusc.2017.08.235
UT WoS 000415219100053
Klíčová slova anglicky Hydrogen soaking; Ni-ion release; NiTi bio-compatibility; NiTi surface passivation; Scanning transmission electron microscopy
Štítky 14110518, EL OK, rivok
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnila: Soňa Böhmová, učo 232884. Změněno: 10. 2. 2019 17:06.
Anotace
The study is focused on Ni-ion release rates from NiTi surfaces exposed in the cell culture media and human vascular endothelial cell (HUVEC) culture environments. The NiTi surface layers situated in the depth of 70 um below a NiTi oxide scale are affected by interactions between the NiTi alloys and the bio-environments. The finding was proved with use of inductively coupled plasma mass spectrometry and electron microscopy experiments. As the exclusive factor controlling the Ni-ion release rates was not only thicknesses of the oxide scale, but also the passivation depth, which was two-fold larger. Our experimental data strongly suggested that some other factors, in addition to the Ni concentration in the oxide scale, admittedly hydrogen soaking deep below the oxide scale, must be taken into account in order to rationalize the concentrations of Ni-ions released into the bio-environments. The suggested role of hydrogen as the surface passivation agent is also in line with the fact that the Ni-ion release rates considerably decrease in NiTi samples that were annealed in controlled hydrogen atmospheres prior to bio-environmental exposures.
Návaznosti
GA15-16336S, projekt VaVNázev: Intersticiální příměsi v tvarově-paměťových slitinách na bázi NiTi
Investor: Grantová agentura ČR, Intersticiální příměsi v tvarově-paměťových slitinách na bázi NiTi
VytisknoutZobrazeno: 23. 4. 2024 16:59