BEŇOVSKÁ, Miroslava, Ondřej WIEWIORKA and Jana PINKAVOVA. Evaluation of FUS-2000 urine analyzer: analytical properties and particle recognition. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION. ABINGDON: INFORMA HEALTHCARE, 2018, vol. 78, 1-2, p. 143-148. ISSN 0036-5513. Available from: https://dx.doi.org/10.1080/00365513.2017.1423108.
Other formats:   BibTeX LaTeX RIS
Basic information
Original name Evaluation of FUS-2000 urine analyzer: analytical properties and particle recognition
Authors BEŇOVSKÁ, Miroslava (203 Czech Republic, guarantor, belonging to the institution), Ondřej WIEWIORKA (203 Czech Republic, belonging to the institution) and Jana PINKAVOVA (203 Czech Republic).
Edition SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, ABINGDON, INFORMA HEALTHCARE, 2018, 0036-5513.
Other information
Original language English
Type of outcome Article in a journal
Field of Study 30100 3.1 Basic medicine
Country of publisher United Kingdom of Great Britain and Northern Ireland
Confidentiality degree is not subject to a state or trade secret
Impact factor Impact factor: 1.380
RIV identification code RIV/00216224:14110/18:00103986
Organization unit Faculty of Medicine
Doi http://dx.doi.org/10.1080/00365513.2017.1423108
UT WoS 000424952600023
Keywords in English Laboratory routines and automation; urinalysis; analytical performance; laboratory statistics; microscopy; medical biochemistry
Tags 14110512, 14110616, rivok
Tags International impact, Reviewed
Changed by Changed by: Soňa Böhmová, učo 232884. Changed: 9/2/2019 22:55.
Abstract
This study evaluates the performance of microscopic part of a hybrid analyzer FUS-2000 (Dirui Industrial Co., Changchun, China), its analytical properties and particle recognition. The evaluation of trueness, repeatability, detection limit, carry-over, linearity range and analytical stability was performed according to Dirui protocol guidelines designed by Dirui Company to guarantee the quality of the instrument. Trueness for low, medium and high-value concentrations was calculated with bias of 15.5, 4.7 and -6.6%, respectively. Detection limit of 5 Ery/l was confirmed. Coefficient of variation of 11.0, 5.2 and 3.8% was measured for within-run repeatability of low, medium and high concentration. Between-run repeatability for daily quality control had coefficient of variation of 3.0%. Carry-over did not exceed 0.05%. Linearity was confirmed for range of 0-16,000 particles/l (R-2=0.9997). The analytical stability had coefficient of variation of 4.3%. Out of 1258 analyzed urine samples, 362 positive were subjected to light microscopy urine sediment analysis and compared to the analyzer results. Cohen's kappa coefficients were calculated to express the concordance. Squared kappa coefficient was 0.927 (red blood cells), 0.888 (white blood cells), 0.908 (squamous epithelia), 0.634 (transitional epithelia), 0.628 (hyaline casts), 0.843 (granular casts) and 0.623 (bacteria). Single kappa coefficients were 0.885 (yeasts) and 0.756 (crystals), respectively. Aforementioned results show good analytical performance of the analyzer and tight agreement with light microscopy of urine sediment.