J 2019

Laser-induced breakdown spectroscopy as a novel readout method for nanoparticle-based immunoassays

MODLITBOVÁ, Pavlína, Zdeněk FARKA, Matěj PASTUCHA, Pavel POŘÍZKA, Karel NOVOTNÝ et. al.

Základní údaje

Originální název

Laser-induced breakdown spectroscopy as a novel readout method for nanoparticle-based immunoassays

Autoři

MODLITBOVÁ, Pavlína (203 Česká republika), Zdeněk FARKA (203 Česká republika, domácí), Matěj PASTUCHA (203 Česká republika, domácí), Pavel POŘÍZKA (203 Česká republika), Karel NOVOTNÝ (203 Česká republika, domácí), Petr SKLÁDAL (203 Česká republika, domácí) a Jozef KAISER (203 Česká republika, garant)

Vydání

Microchimica Acta, Vienna, Springer, 2019, 0026-3672

Další údaje

Jazyk

angličtina

Typ výsledku

Článek v odborném periodiku

Obor

10406 Analytical chemistry

Stát vydavatele

Rakousko

Utajení

není předmětem státního či obchodního tajemství

Odkazy

URL

Impakt faktor

Impact factor: 6.232

Kód RIV

RIV/00216224:14740/19:00110447

Organizační jednotka

Středoevropský technologický institut

DOI

http://dx.doi.org/10.1007/s00604-019-3742-9

UT WoS

000481742100001

Klíčová slova anglicky

Collinear plasma collection; Gold nanoparticles; Laser ablation; Microtiter plate; Sandwich immunoassay; Silver nanoparticles; Streptavidin; Tag-LIBS

Štítky

CF NANO, rivok

Příznaky

Mezinárodní význam, Recenzováno
Změněno: 2. 11. 2024 19:24, Mgr. Adéla Pešková

Anotace

V originále

Laser-induced breakdown spectroscopy (LIBS) was examined as a novel method for readout of microtiter plate immunoassays involving nanoparticles (NP). The so-called Tag-LIBS technique is a sensitive method for the detection of specific biomarkers. It was applied to the determination of NP labels using nanosecond ablation sampling. The NP labels were examined from the bottom of a standard 96-well microtiter plate. Thanks to the flexibility of LIBS instrumentation, both the plasma emission collection and the focusing optics arrangements can be collinearly arranged. The experiments showed that silver NPs and gold NPs can be readily quantified on the bottom of the microtiter plate. Utilizing this technique, a sandwich immunoassay for human serum albumin using streptavidin-coated AgNP labels was developed. The assay has a 10 ng·mL-1 detection limit which is comparable to the sensitivity of fluorometric readout. The main advantage of this LIBS technique is its wide scope in which it enables a detection of almost any type of NP labels, irrespective to any fluorescence or catalytic properties. Owing to the immediate signal response, the relatively simple instrumentation also enables assay automation. The LIBS capability of multi-elemental analyses makes it a promising and fast alternative to other readout techniques, in particular with respect to multiplexed detection of biomarkers.

Návaznosti

CZ.02.1.01/0.0/0.0/16_013/0001728, interní kód MU
Název: CEITEC Nano+ (Akronym: CEITEC Nano)
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, CEITEC Nano+, PO 1 Posilování kapacit pro kvalitní výzkum
LM2015043, projekt VaV
Název: Česká infrastruktura pro integrativní strukturní biologii (Akronym: CIISB)
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, Czech Infrastructure for Integrative Structural Biology
LQ1601, projekt VaV
Název: CEITEC 2020 (Akronym: CEITEC2020)
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, CEITEC 2020
Zobrazeno: 3. 11. 2024 12:39