J 2019

Simultaneous determination of cadmium and iron in different kinds of cereal flakes using high-resolution continuum source atomic absorption spectrometry

ZVĚŘINA, Ondřej, Jan KUTA, Pavel COUFALÍK, Pavlína KOSEČKOVÁ, Josef KOMÁREK et. al.

Basic information

Original name

Simultaneous determination of cadmium and iron in different kinds of cereal flakes using high-resolution continuum source atomic absorption spectrometry

Authors

ZVĚŘINA, Ondřej (203 Czech Republic, guarantor, belonging to the institution), Jan KUTA (203 Czech Republic, belonging to the institution), Pavel COUFALÍK (203 Czech Republic), Pavlína KOSEČKOVÁ (203 Czech Republic, belonging to the institution) and Josef KOMÁREK (203 Czech Republic, belonging to the institution)

Edition

Food Chemistry, Oxford, UK, Elsevier Science, 2019, 0308-8146

Other information

Language

English

Type of outcome

Článek v odborném periodiku

Field of Study

10406 Analytical chemistry

Country of publisher

United Kingdom of Great Britain and Northern Ireland

Confidentiality degree

není předmětem státního či obchodního tajemství

References:

Impact factor

Impact factor: 6.306

RIV identification code

RIV/00216224:14110/19:00110470

Organization unit

Faculty of Medicine

UT WoS

000477667200062

Keywords in English

Cereals; Cadmium; Iron; Simultaneous determination; HR CS GFAAS

Tags

International impact, Reviewed
Změněno: 14/2/2023 13:55, Mgr. Marie Šípková, DiS.

Abstract

V originále

A method for simultaneous determination of cadmium and iron in cereal flakes using high-resolution continuum source graphite furnace atomic absorption spectrometry is presented. Sample digest is introduced into the graphite furnace together with Pd/Mg(NO3)(2) modifier. The primary absorption line of cadmium and adjacent secondary line of iron are used for the determination. Atomization is performed as a two-step process in order to meet ideal conditions for both elements. Interference produced by molecular absorption of PO molecular bands is suppressed by correction model using least squares background correction. Using the proposed method, levels of cadmium and iron were determined in different kinds of cereal flakes, where both elements are of great interest. Working range (0.01-2 mu g L-1 for Cd and 10-500 mu g L-1 for Fe) was suitable for the determination of analytes in samples. The method is fast, robust, and may be routinely used routinely in the analysis of foodstuffs.

Links

EF16_013/0001761, research and development project
Name: RECETOX RI
LM2015051, research and development project
Name: Centrum pro výzkum toxických látek v prostředí (Acronym: RECETOX RI)
Investor: Ministry of Education, Youth and Sports of the CR
MUNI/A/1278/2018, interní kód MU
Name: Prohlubování znalostí v oblasti zdravotních rizik a benefitů výživy, prostředí a životního stylu
Investor: Masaryk University, Category A
MUNI/A/1288/2017, interní kód MU
Name: Výzkum, vývoj a aplikace v analytické a fyzikální chemii (Acronym: AFCH)
Investor: Masaryk University, Category A
ROZV/23/LF21/2019, interní kód MU
Name: Stopové prvky ve výživě a zdraví člověka
Investor: Ministry of Education, Youth and Sports of the CR, Internal development projects