MARTINEK, R., J. BRABLIK, J. KOLARIK, M. LADROVA, J. NEDOMA, R. JAROS, L. SOUSTEK, R. KAHANKOVA, M. FAJKUS, Lubomír VOJTÍŠEK, P. HANZLIKOVA a P. KRUPA. A Low-Cost System for Seismocardiography-Based Cardiac Triggering: A Practical Solution for Cardiovascular Magnetic Resonance Imaging at 3 Tesla. IEEE Access. PISCATAWAY: IEEE Xplore Digital Library, 2019, roč. 7, č. 2019, s. 118608-118629. ISSN 2169-3536. Dostupné z: https://dx.doi.org/10.1109/ACCESS.2019.2936184.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název A Low-Cost System for Seismocardiography-Based Cardiac Triggering: A Practical Solution for Cardiovascular Magnetic Resonance Imaging at 3 Tesla
Autoři MARTINEK, R., J. BRABLIK, J. KOLARIK, M. LADROVA, J. NEDOMA, R. JAROS, L. SOUSTEK, R. KAHANKOVA, M. FAJKUS, Lubomír VOJTÍŠEK (203 Česká republika, garant, domácí), P. HANZLIKOVA a P. KRUPA.
Vydání IEEE Access, PISCATAWAY, IEEE Xplore Digital Library, 2019, 2169-3536.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 10201 Computer sciences, information science, bioinformatics
Stát vydavatele Spojené státy
Utajení není předmětem státního či obchodního tajemství
WWW URL
Impakt faktor Impact factor: 3.745
Kód RIV RIV/00216224:14740/19:00113289
Organizační jednotka Středoevropský technologický institut
Doi http://dx.doi.org/10.1109/ACCESS.2019.2936184
UT WoS 000498548500001
Klíčová slova anglicky Seismocardiography-based cardiac triggering; seismocardiography (SCG); siemens prisma 3T; monitor patient vital signs during MRI; cardiac magnetic resonance imaging (CMRI)
Štítky CF MAFIL, rivok
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnila: Mgr. Pavla Foltynová, Ph.D., učo 106624. Změněno: 12. 5. 2020 13:35.
Anotace
This study describes a pilot clinical validation of a new low-cost system for the continuous monitoring of the human body's cardiorespiratory activities within the magnetic resonance examination area. This study primarily focuses on monitoring cardiac activity and the related cardiac triggering. The patented system tested by the authors is based on seismocardiography (SCG). The study was conducted on 18 subjects on a Siemens Prisma 3T MR scanner. Standard anatomical and diffusion sequences were used to test cardiac activity monitoring. A wide range of commonly used diagnostic sequences were used to test imaging of the heart by means of cardiac triggering. System functionality was verified against a commercially available electrocardiography (ECG) system. Monitoring of cardiac activity (detection of the R-R interval in ECG and the AO-AO interval in SCG) was objectively evaluated by determining the overall probability of correct detection (ACC), sensitivity (SE), positive predictive value (PPV), and harmonic mean between SE and PPV, i.e. F1. Imaging quality control using Cardiac Triggering was performed by subjective evaluation of images by the physicians. The study conducted clearly confirmed the functionality of the system in terms of continuous cardiac activity monitoring. In all 18 subjects, a mean PPV > 99% was achieved; F1 > 99 %; SE > 99 %; ACC > 98 %; 1.96 sigma < 3.5 bpm. Also, Cardiac Triggering functionality was confirmed by the physicians on the basis of analyzing cardiac images using the T1/T2 balanced echo sequences and the T1 flash sequence measured natively.
Návaznosti
LM2015062, projekt VaVNázev: Národní infrastruktura pro biologické a medicínské zobrazování
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, National research infrastructure for biological and medical imaging
VytisknoutZobrazeno: 25. 4. 2024 17:56