Real-time measurement of ICD lead motion during stereotactic body radiotherapy of ventricular tachycardia

KNYBEL, Lukas, Jakub CVEK, Radek NEUWIRTH, Otakar JIRAVSKÝ, Jan HECKO et. al.

Basic information

Original name

Real-time measurement of ICD lead motion during stereotactic body radiotherapy of ventricular tachycardia

Authors

KNYBEL, Lukas (203 Czech Republic), Jakub CVEK (203 Czech Republic, guarantor), Radek NEUWIRTH (203 Czech Republic, belonging to the institution), Otakar JIRAVSKÝ (203 Czech Republic), Jan HECKO (203 Czech Republic), Marek PENHAKER (203 Czech Republic), Marek SRAMKO (203 Czech Republic) and Josef KAUTZNER (203 Czech Republic)

Edition

Reports of Practical Oncology and Radiotherapy, GDANSK, VIA MEDICA, 2021, 1507-1367

Other information

Language

English

Type of outcome

Článek v odborném periodiku

Field of Study

30224 Radiology, nuclear medicine and medical imaging

Country of publisher

Poland

Confidentiality degree

není předmětem státního či obchodního tajemství

References:

RIV identification code

RIV/00216224:14110/21:00121087

Organization unit

Faculty of Medicine

UT WoS

000641613400016

Keywords in English

ventricular tachycardia; stereotactic body radiotherapy

Tags

Tags

International impact, Reviewed
Změněno: 21/4/2022 10:19, Mgr. Tereza Miškechová

Abstract

V originále

Background: Here we aimed to evaluate the respiratory and cardiac-induced motion of a ICD lead used as surrogate in the heart during stereotactic body radiotherapy (SBRT) of ventricular tachycardia (VT). Data provides insight regarding motion and motion variations during treatment. Materials and methods: We analyzed the log files of surrogate motion during SBRT of ventricular tachycardia performed in 20 patients. Evaluated parameters included the ICD lead motion amplitudes; intrafraction amplitude variability; correlation error between the ICD lead and external markers; and margin expansion in the superior-inferior (SI), latero-lateral (LL), and anterior-posterior (AP) directions to cover 90% or 95% of all amplitudes. Results: In the SI, LL, and AP directions, respectively, the mean motion amplitudes were 5.0 +/- 2.6, 3.4. +/- 1.9, and 3.1 +/- 1.6 mm. The mean intrafraction amplitude variability was 2.6 +/- 0.9, 1.9 +/- 1.3, and 1.6 +/- 0.8 mm in the SI, LL, and AP directions, respectively. The margins required to cover 95% of ICD lead motion amplitudes were 9.5, 6.7, and 5.5 mm in the SI, LL, and AP directions, respectively. The mean correlation error was 2.2 +/- 0.9 mm. Conclusions: Data from online tracking indicated motion irregularities and correlation errors, necessitating an increased CTV-PTV margin of 3 mm. In 35% of cases, the motion variability exceeded 3 mm in one or more directions. We recommend verifying the correlation between CTV and surrogate individually for every patient, especially for targets with posterobasal localization where we observed the highest difference between the lead and CTV motion.