Autonomous capillary electrophoresis processing and analysis of dried blood spots for high-throughput determination of uric acid
MORAVČÍK, Ondrej, Miloš DVOŘÁK and Pavel KUBÁŇBasic information
Original name
Autonomous capillary electrophoresis processing and analysis of dried blood spots for high-throughput determination of uric acid
Authors
MORAVČÍK, Ondrej (703 Slovakia, belonging to the institution), Miloš DVOŘÁK and Pavel KUBÁŇ (guarantor)
Edition
Analytica Chimica Acta, Amsterdam, Elsevier, 2023, 0003-2670
Other information
Language
English
Type of outcome
Článek v odborném periodiku
Field of Study
10406 Analytical chemistry
Country of publisher
Netherlands
Confidentiality degree
není předmětem státního či obchodního tajemství
References:
Impact factor
Impact factor: 6.200 in 2022
RIV identification code
RIV/00216224:14310/23:00131620
Organization unit
Faculty of Science
UT WoS
001002501500001
Keywords in English
Automation; Dried blood spot; Capillary electrophoresis; Uric acid
Tags
Tags
International impact, Reviewed
Změněno: 7/9/2023 10:38, Mgr. Marie Šípková, DiS.
Abstract
V originále
A new set-up for fully autonomous and high-throughput capillary electrophoresis (CE) analyses of dried blood spot (DBS) samples is presented. The DBS samples were prepared by collecting exactly 5 μL of capillary blood from a finger-prick onto a pre-punched DBS disc in a disposable plastic CE vial and by in-vial blood drying. The vials with the DBS samples were then loaded into a commercial CE instrument for a fully unmanned sample processing and analysis. A fused-silica capillary of the CE instrument was first used for the transfer of 100 μL of elution solvent to each vial, in-vial DBS elution, and in-vial eluate homogenization. The same capillary was also used for at-line injection, separation, and selective analysis of the resulting eluates. Novel CE sequences were tailor-programmed for consecutive processing and analyses of multiple DBSs, which facilitated a fully autonomous determination of uric acid with a throughput of 240 DBS samples per day (24 h). The presented analytical protocol (using 100 μm i. d./30 cm capillary; 30 mM 2-(N-morpholino)-ethanesulfonic acid, 30 mM l-histidine, and 30 μM cetyltrimethylammonium bromide background electrolyte solution; and UV detection at 292 nm) provided excellent precision at endogenous and spiked uric acid concentrations with RSD values of peak areas below 3.2%. Calibration curves were linear over the 33.3 − 1200 μM range (R2 better than 0.998), limits of detection and quantification in the original capillary blood were 10 and 33.3 μM, respectively, and were well below the uric acid clinical range (140–420 μM). The stability of uric acid in DBS samples stored at laboratory temperature for up to 2 months was also excellent demonstrating less than a 4.2% decrease in uric acid concentrations. The actual set-up might thus be highly attractive for clinical subjects and laboratories because it is minimally invasive and requires minimum intervention from laboratory staff.