Development of a combined setup for simultaneous detection of total and glycated haemoglobin content in blood samples

PŘIBYL, Jan and Petr SKLÁDAL

Basic information

Original name

Development of a combined setup for simultaneous detection of total and glycated haemoglobin content in blood samples

Name in Czech

Vývoj kombinovaného systému pro současné stanovení celkového a glykovaného hemoglobinu ve vzorcích krve

Authors

PŘIBYL, Jan (203 Czech Republic) and Petr SKLÁDAL (203 Czech Republic, guarantor)

Edition

Biosensors & Bioelectronics, Oxford, Elsevier Advanced Technology, 2006, 0956-5663

Other information

Language

English

Type of outcome

Článek v odborném periodiku

Field of Study

10600 1.6 Biological sciences

Country of publisher

United Kingdom of Great Britain and Northern Ireland

Confidentiality degree

není předmětem státního či obchodního tajemství

Impact factor

Impact factor: 4.132

RIV identification code

RIV/00216224:14310/06:00016801

Organization unit

Faculty of Science

UT WoS

000236756100015

Keywords in English

Biosensor; Piezoelectric sensor; Quartz crystal microbalance; Flow injection analysis

Tags

International impact, Reviewed
Změněno: 16/6/2007 22:32, prof. RNDr. Petr Skládal, CSc.

Abstract

V originále

An original setup for analysis of glycated haemoglobin (HbA,) in blood is reported. The construction employed a combination of the piezoelectric biosensor for glycated haemoglobin and the flow-through photometric sensor for total haemoglobin (Hb). The modification of gold electrodes with 3-aminophenylboronic acid (APBA) as a specific ligand was studied; the chemisorbed conjugate of APBA with a long-chain thiocompound provided the best affinity for Hb(A1c). The effect of various operating parameters, such as flow rate and instrumental setup, was optimised. The total haemoglobin content was analysed as absorbance of the haemoglobin-cyanide derivative at 540 nm. Only one standard (calibrator) diluted in various ratio was necessary for calibration and 1 mu l of blood was sufficient for analysis. The full range of Hb(A1c) content (4-15%) in blood can be analysed; the working ranges of total and glycated haemoglobin were 50-2000 and 10-90 mu g/ml, respectively. The developed method was successfully evaluated on blood samples collected from diabetics.

In Czech

Byl vyvinut kombinovaný systém na bázi piezoelektrického biosensoru s imobilizovanou kyselinou aminofenylboritou pro specifickou detekci glykvaného hemoglobinu, paralelně se spektrofotometricky stanovuje celkový hemoglobin. Systém je spojen s automatický průtočným analyzátorem FIAlab a je vhodný pro měření vzorků krve.