To study the microvessels in bulbar conjunctiva, we conducted an experiment in goat with a pneumatically driven left heart bypass pump, which was replaced with an undulation pump-left ventricle assist device for 9 days. Three flow patterns were tested: complete pulsatile, continuous, and percentage of pulsatile. We studied the morphology of arterioles and venules of the bulbar conjunctiva using photograph records. The setting up of continuous flow caused global vasoconstriction (significant in venules - P 0.05). During the pumping in the pulsatile and percentage of pulsatile modes, no significant changes of microvessel morphology were observed. The findings described could point to some disturbances in the microcirculatory bed in conditions of continuous flow.
Česky
Mikrocirkulace na oční spojivce byla studována v experimentu u kozla, u kterého byla pneumatická bypassová pumpa nahrazující levé srdce zaměněna na 9 dní za undulační pumpu - podpůrné zařízení levé komory. Byly zkoumány tři způsoby nastavení průtoku: čistě pulzační, kontinuální a procentuálně pulzační. Studovali jsme morfologii arteriol a venul na oční spojivce pomocí fotografických záznamů. Kontinuální průtok na pumpě způsobil celkovou vazokonstrikci (signifikantní ve venulách - P 0.05). Během průtoku ve ostatních dvou módech nebyly pozorovány žádné signifikantní změny mikrocirkulace - cévní morfologie. Naše nálezy by mohly vysvětlit určité poruchy v mikrocirkulaci během nastavení pumpy na kontinuální průtok.