J 2006

Influence of flow design on microcirculation in conditions of undulation pump-left ventricle assist device testing

DOBŠÁK, Petr, Marie NOVÁKOVÁ, Atsushi BABA, Jaromír VAŠKŮ, Takashi ISOYAMA et. al.

Basic information

Original name

Influence of flow design on microcirculation in conditions of undulation pump-left ventricle assist device testing

Name in Czech

Vliv nastavení způsobu průtoku na mikrocirkulaci u undulační pumpy - levokomorového podpůrného zařízení

Authors

DOBŠÁK, Petr (203 Czech Republic, guarantor), Marie NOVÁKOVÁ (203 Czech Republic), Atsushi BABA (392 Japan), Jaromír VAŠKŮ (203 Czech Republic), Takashi ISOYAMA (392 Japan), Itsuro SAITO (392 Japan), Yusuke ABE (392 Japan), Tsuneo CHINZEI (392 Japan) and Kou IMACHI (392 Japan)

Edition

Artificial Organs, Merilend, USA, Blackwell Science, Inc. 2006, 0160-564X

Other information

Language

English

Type of outcome

Článek v odborném periodiku

Field of Study

30105 Physiology

Country of publisher

United States of America

Confidentiality degree

není předmětem státního či obchodního tajemství

Impact factor

Impact factor: 1.903

Organization unit

Faculty of Medicine

UT WoS

000237811700010

Keywords in English

mechanical cardiac support;pulsatile and nonpulsatile flow;microcirculation;conjunctival angioscopy

Tags

International impact, Reviewed
Změněno: 19/6/2009 16:31, prof. MUDr. Marie Nováková, Ph.D.

Abstract

V originále

To study the microvessels in bulbar conjunctiva, we conducted an experiment in goat with a pneumatically driven left heart bypass pump, which was replaced with an undulation pump-left ventricle assist device for 9 days. Three flow patterns were tested: complete pulsatile, continuous, and percentage of pulsatile. We studied the morphology of arterioles and venules of the bulbar conjunctiva using photograph records. The setting up of continuous flow caused global vasoconstriction (significant in venules - P 0.05). During the pumping in the pulsatile and percentage of pulsatile modes, no significant changes of microvessel morphology were observed. The findings described could point to some disturbances in the microcirculatory bed in conditions of continuous flow.

In Czech

Mikrocirkulace na oční spojivce byla studována v experimentu u kozla, u kterého byla pneumatická bypassová pumpa nahrazující levé srdce zaměněna na 9 dní za undulační pumpu - podpůrné zařízení levé komory. Byly zkoumány tři způsoby nastavení průtoku: čistě pulzační, kontinuální a procentuálně pulzační. Studovali jsme morfologii arteriol a venul na oční spojivce pomocí fotografických záznamů. Kontinuální průtok na pumpě způsobil celkovou vazokonstrikci (signifikantní ve venulách - P 0.05). Během průtoku ve ostatních dvou módech nebyly pozorovány žádné signifikantní změny mikrocirkulace - cévní morfologie. Naše nálezy by mohly vysvětlit určité poruchy v mikrocirkulaci během nastavení pumpy na kontinuální průtok.