Analysis of brominated flame retardants and their derivatives by atmospheric pressure chemical ionization using gas chromatography coupled to tandem quadrupole mass spectrometry
GENG, Dawei, Petr KUKUČKA a Ingrid Ericson JOGSTENZákladní údaje
Originální název
Analysis of brominated flame retardants and their derivatives by atmospheric pressure chemical ionization using gas chromatography coupled to tandem quadrupole mass spectrometry
Autoři
GENG, Dawei (156 Čína), Petr KUKUČKA (203 Česká republika, garant, domácí) a Ingrid Ericson JOGSTEN (752 Švédsko)
Vydání
Talanta, AMSTERDAM, Elsevier, 2017, 0039-9140
Další údaje
Jazyk
angličtina
Typ výsledku
Článek v odborném periodiku
Obor
10406 Analytical chemistry
Stát vydavatele
Nizozemské království
Utajení
není předmětem státního či obchodního tajemství
Odkazy
Impakt faktor
Impact factor: 4.244
Kód RIV
RIV/00216224:14310/17:00096416
Organizační jednotka
Přírodovědecká fakulta
UT WoS
000389088700083
Klíčová slova anglicky
BFR; PBDE; MeO-PBDE; gas chromatography-atmospheric pressure chemical ionization (GC-APCI); Tandem quadrupole mass spectrometry (MS/MS)
Příznaky
Mezinárodní význam, Recenzováno
Změněno: 28. 3. 2018 14:47, Ing. Nicole Zrilić
Anotace
V originále
A validated method using an atmospheric pressure chemical ionization source for coupling gas chromatography (GC-APCI) to tandem quadrupole mass spectrometry (MS/MS) for the determination of brominated flame retardahts (BFRs) is presented. Polybrominated diphenyl ethers (PBDEs), their methoxylated derivatives (MeO-PBDEs) and other emerging BFRs were included in this study. The method showed good linearity and repeatability. The relative standard deviation (RSD) of the relative response factors (RRFs) of all compounds was below 16%. Repeatability for BFRs was tested on one or two concentration levels of calibration standards with RSDs for RRFs below 16%. The lowest calibration standards (0.075-0.1 pg/mu L for emerging BFRs, BDE 209 and MeO-PBDEs mixtures, 0.625-6.25 pg/mu L for Br1-9 PBDEs mixtures) were used as instrument detection limits (IDL). The method was applied on biotic samples, including fish, osprey, and seal. In general, BDE 209 and decabromodiphenyl ethane (DBDPE) were detected in 50% of the seal samples. A 100% detection rate was achieved for 6-MeO-BDE 47 in all the samples (72-580 pg/g ww in osprey samples, 24,00096,000 pg/g ww in seal samples and 78-99 pg/g ww in fish samples). All Br3-6 PBDEs (BDE 28, 47, 99, 100, 153, 154) were detected in all the samples (ranging from 12 to 20,000 pg/g ww), while BDE 183 was detected in 60% of the osprey eggs, 20% of the seal samples and below MDL in all fish samples. The results presented indicate the capability of the GC-APCI-MS/MS system for the detection and quantification of BFRs.
Návaznosti
LM2015051, projekt VaV |
| ||
LO1214, projekt VaV |
|