Detailed Information on Publication Record
2016
Neutron/gamma-ray discrimination through measures of fit
AMIRI, Moslem, Václav PŘENOSIL and František CVACHOVECBasic information
Original name
Neutron/gamma-ray discrimination through measures of fit
Name in Czech
Separace neutronů a fama záření metodou FITt
Authors
AMIRI, Moslem (364 Islamic Republic of Iran, guarantor, belonging to the institution), Václav PŘENOSIL (203 Czech Republic, belonging to the institution) and František CVACHOVEC (203 Czech Republic, belonging to the institution)
Edition
Lisbon, 2015 4th International Conference on Advancements in Nuclear Instrumentation Measurement Methods and their Applications (ANIMMA), p. 1-5, 5 pp. 2016
Publisher
IEEE
Other information
Language
English
Type of outcome
Stať ve sborníku
Field of Study
20206 Computer hardware and architecture
Confidentiality degree
není předmětem státního či obchodního tajemství
Publication form
electronic version available online
RIV identification code
RIV/00216224:14330/16:00087671
Organization unit
Faculty of Informatics
ISBN
978-1-4799-9918-7
UT WoS
000398710600070
Keywords (in Czech)
organická scintilátor, čítač; statistický test; metoda FIT; neutronová spektroskopie
Keywords in English
organic scintillator; counter; statistical test; measures of fit; neutron spectroscopy
Tags
International impact, Reviewed
Změněno: 13/5/2020 20:55, RNDr. Pavel Šmerk, Ph.D.
V originále
Statistical tests and their underlying measures of fit can be utilized to separate neutron/gamma-ray pulses in a mixed radiation field. In this article, first the application of a sample statistical test is explained. Fit measurement-based methods require true pulse shapes to be used as reference for discrimination. This requirement makes practical implementation of these methods difficult; typically another discrimination approach should be employed to capture samples of neutrons and gamma-rays before running the fit-based technique. In this article, we also propose a technique to eliminate this requirement. These approaches are applied to several sets of mixed neutron and gamma-ray pulses obtained through different digitizers using stilbene scintillator in order to analyze them and measure their discrimination quality.
In Czech
Statistické testy a jejich základní postupy mohou být použity pro separaci neutronů/gama záření ve smíšeném radiačním poli. V tomto článku je popsána první aplikace vzorku statistického testu. FIT metoda měření vyžaduje skutečné tvary impulsů, které mají být použity jako referenční pro vlastní diskriminaci. Tento požadavek metody je pro praktickou aplikaci problematický - obvykle by měla být použita jiný postup diskriminace pro zachycování vzorků neutronů a gama záření před aplikací metody FIT. V tomto článku se také navrhujeme techniku k odstranění tohoto požadavku. Tyto přístupy jsou aplikovány na několika sadách vzorků smíšeného pole neutronů a gama záření získaných několika typy digitizérů s použití scintiátoru stilbene pro získání představy o jejich podílu na kvalitě diskriminace.
Links
TA01011383, research and development project |
|