2017
Quantification of elemental mapping of heterogeneous geological sample by laser ablation inductively coupled plasma mass spectrometry
VACULOVIČ, Tomáš, Karel BREITER, Zuzana KORBELOVÁ, Natálie VENCLOVÁ, Kateřina TOMKOVÁ et. al.Základní údaje
Originální název
Quantification of elemental mapping of heterogeneous geological sample by laser ablation inductively coupled plasma mass spectrometry
Autoři
VACULOVIČ, Tomáš (203 Česká republika, garant, domácí), Karel BREITER (203 Česká republika), Zuzana KORBELOVÁ (203 Česká republika), Natálie VENCLOVÁ (203 Česká republika), Kateřina TOMKOVÁ (203 Česká republika), Šárka JONÁŠOVÁ (203 Česká republika) a Viktor KANICKÝ (203 Česká republika, domácí)
Vydání
Microchemical Journal, Amsterdam, Elsevier Science, 2017, 0026-265X
Další údaje
Jazyk
angličtina
Typ výsledku
Článek v odborném periodiku
Obor
10406 Analytical chemistry
Stát vydavatele
Česká republika
Utajení
není předmětem státního či obchodního tajemství
Impakt faktor
Impact factor: 2.746
Kód RIV
RIV/00216224:14310/17:00094928
Organizační jednotka
Přírodovědecká fakulta
UT WoS
000403121900029
Klíčová slova anglicky
LA-ICP-MS; quantification; imaging; archaeological samples; geological samples
Změněno: 11. 4. 2018 10:36, Ing. Nicole Zrilić
Anotace
V originále
Quantification procedures based on the external calibration with total sum content standardization (ECTSCS) and the external calibration with internal standardization (ECIS) were compared in the analysis of glass materials by laser ablation inductively coupled plasma spectrometry (LA-ICP-MS) with a 213 nm Nd:YAG laser ablation device and a quadrupole mass spectrometer. Silicon was used as the internal reference element for ECIS. Applicability of these approaches was statistically tested using both homogeneous and inhomogeneous samples. The former type was represented by the standard reference material NIST 612 and an ancient glass bead (Early Bronze Age), while the latter was a mediaeval glass artefact (8th - 9th century). Homogeneity of samples was examined by EMP analysis. Identical results were obtained by both calibration methods for homogeneous samples, while underestimated contents resulted from quantification with ECIS method in comparison to EMP analyses of the same target area of inhomogeneous glass. Finally, elemental maps of zonal muscovite grain were recorded and quantified using ECTSCS method and verified by EMP analysis.
Návaznosti
GA14-13600S, projekt VaV |
| ||