2017
Non-Invasive Fetal Monitoring: A Maternal Surface ECG Electrode Placement-Based Novel Approach for Optimization of Adaptive Filter Control Parameters Using the LMS and RLS Algorithms
MARTINEK, Radek, Radana KAHANKOVA, Homer NAZERAN, Jaromir KONECNY, Janusz JEZEWSKI et. al.Základní údaje
Originální název
Non-Invasive Fetal Monitoring: A Maternal Surface ECG Electrode Placement-Based Novel Approach for Optimization of Adaptive Filter Control Parameters Using the LMS and RLS Algorithms
Autoři
MARTINEK, Radek (203 Česká republika), Radana KAHANKOVA (203 Česká republika), Homer NAZERAN (840 Spojené státy), Jaromir KONECNY (203 Česká republika), Janusz JEZEWSKI (616 Polsko), Petr JANKŮ (203 Česká republika, garant, domácí), Petr BILIK (203 Česká republika), Jan ZIDEK (203 Česká republika), Jan NEDOMA (203 Česká republika) a Marcel FAJKUS (203 Česká republika)
Vydání
Sensors, BASEL, MDPI AG, 2017, 1424-8220
Další údaje
Jazyk
angličtina
Typ výsledku
Článek v odborném periodiku
Obor
10405 Electrochemistry
Stát vydavatele
Švýcarsko
Utajení
není předmětem státního či obchodního tajemství
Impakt faktor
Impact factor: 2.475
Kód RIV
RIV/00216224:14110/17:00100145
Organizační jednotka
Lékařská fakulta
UT WoS
000404553300215
Klíčová slova anglicky
fetal ECG; adaptive filtering; Least Mean Squares (LMS) algorithm; Recursive Least Squares (RLS) algorithm
Štítky
Příznaky
Mezinárodní význam, Recenzováno
Změněno: 2. 3. 2018 16:27, Soňa Böhmová
Anotace
V originále
This paper is focused on the design, implementation and verification of a novel method for the optimization of the control parameters (such as step size mu and filter order N) of LMS and RLS adaptive filters used for noninvasive fetal monitoring. The optimization algorithm is driven by considering the ECG electrode positions on the maternal body surface in improving the performance of these adaptive filters. The main criterion for optimal parameter selection was the Signal-to-Noise Ratio (SNR). We conducted experiments using signals supplied by the latest version of our LabVIEW-Based Multi-Channel Non-Invasive Abdominal Maternal-Fetal Electrocardiogram Signal Generator, which provides the flexibility and capability of modeling the principal distribution of maternal/fetal ECGs in the human body. Our novel algorithm enabled us to find the optimal settings of the adaptive filters based on maternal surface ECG electrode placements. The experimental results further confirmed the theoretical assumption that the optimal settings of these adaptive filters are dependent on the ECG electrode positions on the maternal body, and therefore, we were able to achieve far better results than without the use of optimization. These improvements in turn could lead to a more accurate detection of fetal hypoxia. Consequently, our approach could offer the potential to be used in clinical practice to establish recommendations for standard electrode placement and find the optimal adaptive filter settings for extracting high quality fetal ECG signals for further processing. Ultimately, diagnostic-grade fetal ECG signals would ensure the reliable detection of fetal hypoxia.