MAKHNEVA, Ekaterina, Zdeněk FARKA, Petr SKLÁDAL a Lenka ZAJÍČKOVÁ. Cyclopropylamine plasma polymer surfaces for label-free SPR and QCM immunosensing of Salmonella. Sensors and Actuators B: Chemical. Amsterdam: Elsevier B.V., 2018, roč. 276, DEC, s. 447-455. ISSN 0925-4005. Dostupné z: https://dx.doi.org/10.1016/j.snb.2018.08.055.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Cyclopropylamine plasma polymer surfaces for label-free SPR and QCM immunosensing of Salmonella
Autoři MAKHNEVA, Ekaterina (643 Rusko, domácí), Zdeněk FARKA (203 Česká republika, domácí), Petr SKLÁDAL (203 Česká republika, domácí) a Lenka ZAJÍČKOVÁ (203 Česká republika, garant, domácí).
Vydání Sensors and Actuators B: Chemical, Amsterdam, Elsevier B.V. 2018, 0925-4005.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 10406 Analytical chemistry
Stát vydavatele Švýcarsko
Utajení není předmětem státního či obchodního tajemství
WWW URL
Impakt faktor Impact factor: 6.393
Kód RIV RIV/00216224:14740/18:00101169
Organizační jednotka Středoevropský technologický institut
Doi http://dx.doi.org/10.1016/j.snb.2018.08.055
UT WoS 000444637600056
Klíčová slova anglicky Plasma polymerization; Amine films; Surface plasmon resonance; Quartz crystal microbalance; Immunosensor; Label-free detection
Štítky CF NANO, rivok
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnila: Mgr. Pavla Foltynová, Ph.D., učo 106624. Změněno: 23. 4. 2019 09:00.
Anotace
We report on the first successful application of nitrogen-containing functional surfaces prepared by plasma polymerization in surface plasmon resonance (SPR) immunosensing. The plasma polymers (PPs) were deposited from cyclopropylamine (CPA) vapors onto gold surfaces of SPR and quartz crystal microbalance (QCM) sensors. It provided an excellent platform for a stable immobilization of antibodies using glutaraldehyde (GA) activation. The performance of the SPR immunosensor was proven by a model pair of monoclonal antibody (Ab) AL-01 and human serum albumin as well as by the selective detection of bacterial pathogen Salmonella Typhimurium using the immobilized anti-Salmonella Ab. A baseline drift of the measured SPR signal, caused by the long-term reactivity of CPA plasma polymers, was well stabilized by 18 h immersion in phosphate buffered saline prior to the GA activation. The immunosensors can be successfully regenerated several times using 10 mM NaOH. The limit of detection of 10^5 CFU/mL and a wide linear response were achieved for Salmonella. The assay parameters are comparable to the conventional label-free methods. However, the replacement of conventional immobilization matrix with the CPA PPs offers an advantage of excellent adhesion to gold QCM and SPR surfaces achieved by fast and eco-friendly procedure compatible with the vacuum deposition of gold.
Návaznosti
GA18-12774S, projekt VaVNázev: Plazmové polymery připravené na nanovlákenných membránách pro inženýrství cévní tkáně
Investor: Grantová agentura ČR, Plazmové polymery připravené na nanovlákenných membránách pro inženýrství cévní tkáně
LM2015043, projekt VaVNázev: Česká infrastruktura pro integrativní strukturní biologii (Akronym: CIISB)
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, Czech Infrastructure for Integrative Structural Biology
LQ1601, projekt VaVNázev: CEITEC 2020 (Akronym: CEITEC2020)
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, CEITEC 2020
VytisknoutZobrazeno: 22. 7. 2024 07:18