ANDRAŠINA, Tomáš, Tomáš ROHAN, Tomáš JŮZA, Peter MATKULČÍK, Jana POKORNÁ a Vlastimil VÁLEK. Irreversible electroporation for treatment of metal stent occlusion in biliary tract – ex vivo experimental model. In VÁLEK, Vlastimil. CIRSE 2018. 2018. ISSN 0174-1551. Dostupné z: https://dx.doi.org/10.1007/s00270-018-2021-1.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Irreversible electroporation for treatment of metal stent occlusion in biliary tract – ex vivo experimental model
Autoři ANDRAŠINA, Tomáš (703 Slovensko, domácí), Tomáš ROHAN (203 Česká republika, domácí), Tomáš JŮZA (203 Česká republika, domácí), Peter MATKULČÍK (703 Slovensko, domácí), Jana POKORNÁ (203 Česká republika, domácí) a Vlastimil VÁLEK (203 Česká republika, domácí).
Vydání CIRSE 2018, 2018.
Další údaje
Originální jazyk angličtina
Typ výsledku Konferenční abstrakt
Obor 30224 Radiology, nuclear medicine and medical imaging
Stát vydavatele Česká republika
Utajení není předmětem státního či obchodního tajemství
Impakt faktor Impact factor: 1.928
Kód RIV RIV/00216224:14110/18:00103763
Organizační jednotka Lékařská fakulta
ISSN 0174-1551
Doi http://dx.doi.org/10.1007/s00270-018-2021-1
UT WoS 000446679300001
Klíčová slova anglicky irreversible electroporation; ex vivo experiment; biliary metal stent
Příznaky Mezinárodní význam
Změnil Změnil: RNDr. Pavel Šmerk, Ph.D., učo 3880. Změněno: 31. 5. 2022 12:16.
Anotace
Purpose To prove the safety and feasibility of irreversible electroporation (IRE) using tubular catheters to treat biliary metal stent occlusions in an ex vivo experiment model. Material and methods IRE was performed using 3-electrode tubular IRE catheters placed in metal stents (EGIS biliary stent 10x80 mm) in ex vivo porcine liver models. The IRE catheter was connected to an IRE generator, with two electrodes set as active and one as indifferent. One-hundred 100 us pulses at voltages 300V, 650V, 1000V, and 1300V were used. Stent occlusion was simulated using porcine liver tissues of different thickness (2±1 mm and 5±1 mm). Five scenarios of geometry settings between stent, electrodes, and inner tissue were analysed. Values of electric current, impedance, and power output were measured. Potentially dangerous thermal effects were monitored thermographically and visually on the stent and surrounding parenchyma. Results The IRE procedure was feasible for all settings of the voltages of 300V and 650V. The maximum current limit of the generator was exceeded in case of low profile ingrowth tissue connected with one electrode and in all settings with use of 1300V. In these cases, thermal changes of the liver tissue were visualised. In contrast, heat changes in protocols using 300-650V were recorded only in range of 1,0-4,5°C. Significant difference of impedance between one-electrode and two-electrodes ranges of simulated stent occlusion was observed (p = 0,02, Mann-Whitney). Conclusion IRE using a 3-electrode tubular catheter shows feasibility for treatment of metal stent ingrowth in ex vivo experiments. We have established a safe and applicable IRE protocol for further in vivo experiments.
Návaznosti
MUNI/A/1255/2017, interní kód MUNázev: Onkologické radiologické intervence a jejich přínos v rámci komplexní onkologické léčby, srovnání s celorepublikovými výsledky onkologické léčby vybraných diagnóz
Investor: Masarykova univerzita, Onkologické radiologické intervence a jejich přínos v rámci komplexní onkologické léčby, srovnání s celorepublikovými výsledky onkologické léčby vybraných diagnóz, DO R. 2020_Kategorie A - Specifický výzkum - Studentské výzkumné projekty
NV15-32484A, projekt VaVNázev: Využití nových biotechnologií v prevenci a léčbě stenóz žlučových cest
ROZV/24/LF/2018, interní kód MUNázev: LF - Příspěvek na IP 2108
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, LF - Příspěvek na IP 2108, Interní rozvojové projekty
VytisknoutZobrazeno: 13. 9. 2024 13:00