a 2018

Prevention of benign stenosis due to endoluminal irreversible electroporation in bile ducts

ANDRAŠINA, Tomáš, Jiří PÁNEK, Dalibor ČERVINKA, Iva SVOBODOVÁ, Tomáš ROHAN et. al.

Základní údaje

Originální název

Prevention of benign stenosis due to endoluminal irreversible electroporation in bile ducts

Autoři

ANDRAŠINA, Tomáš (703 Slovensko, domácí), Jiří PÁNEK (203 Česká republika, domácí), Dalibor ČERVINKA, Iva SVOBODOVÁ (203 Česká republika), Tomáš ROHAN (203 Česká republika, domácí), Jakub HUSTÝ (840 Spojené státy, domácí), Hiroko MARTIN (840 Spojené státy, domácí) a Vlastimil VÁLEK (203 Česká republika, domácí)

Vydání

CIRSE 2018, 2018

Další údaje

Jazyk

angličtina

Typ výsledku

Konferenční abstrakt

Obor

30224 Radiology, nuclear medicine and medical imaging

Stát vydavatele

Česká republika

Utajení

není předmětem státního či obchodního tajemství

Impakt faktor

Impact factor: 1.928

Kód RIV

RIV/00216224:14110/18:00103798

Organizační jednotka

Lékařská fakulta

ISSN

UT WoS

000446679300001

Klíčová slova anglicky

Irreversible electroporation; biodegradable stent; porcine model

Příznaky

Mezinárodní význam
Změněno: 31. 5. 2022 12:17, RNDr. Pavel Šmerk, Ph.D.

Anotace

V originále

Purpose To evaluate the possibility of preventing the formation of benign stenosis in healthy bile ducts due to endoluminal irreversible electroporation (IRE). Material and methods Endoluminal IRE of the common bile duct was performed on 15 porcine models using an endoluminal device inserted during laparotomy. The IRE procedure was conducted with 90 pulses of 900-1500 V. Models categorized in group A (n=5) underwent insertion of a biodegradable stent (30x10 mm, Ella-CS) in the ablation site. In group B, the ablation site of the models (n=10) remained untreated. All pigs were euthanized 22 days post-procedure. Biochemical liver tests and MRI were employed for follow-up on days 0, 3, and 22 after ablation. Results The IRE treatment and stent implantation was successful in all procedures. All animals survived over the defined study period. All biodegradable stents remained in place of ablation until the time of autopsy. In group A, serum bilirubin levels were slightly elevated immediately after the procedure (average 19.08 µmol/l compared to 8.33 µmol/l in group B). On day 22, significant elevation of serum bilirubin levels and AST levels were observed in group B (mean 1.73 µmol/l and 18.28 µmol/l, 0.79 µkat/l and 1.74 µkat/l in groups A and B respectively, p<0,05). MRCP sequence on day 22 showed mild or severe dilatation of intrahepatic biliary ducts in a majority of the models (n=9, 90%) in group B. By comparison, the image analysis in group A revealed no intrahepatic dilatation. Conclusion The preventive implantation of biodegradable stents following IRE of bile ducts should be a promising technique to deal with benign stenosis formation after treatment.

Návaznosti

MUNI/A/1255/2017, interní kód MU
Název: Onkologické radiologické intervence a jejich přínos v rámci komplexní onkologické léčby, srovnání s celorepublikovými výsledky onkologické léčby vybraných diagnóz
Investor: Masarykova univerzita, Onkologické radiologické intervence a jejich přínos v rámci komplexní onkologické léčby, srovnání s celorepublikovými výsledky onkologické léčby vybraných diagnóz, DO R. 2020_Kategorie A - Specifický výzkum - Studentské výzkumné projekty
NV15-32484A, projekt VaV
Název: Využití nových biotechnologií v prevenci a léčbě stenóz žlučových cest
ROZV/24/LF/2018, interní kód MU
Název: LF - Příspěvek na IP 2108
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, LF - Příspěvek na IP 2108, Interní rozvojové projekty