FILIPOVSKÝ, Jan, Jitka SEIDLEROVÁ, Jiří CERAL, Petra VYSOČANOVÁ, Jiří ŠPÁC, Miroslav SOUČEK, Ivan ŘIHÁČEK, Markéta MATEŘÁNKOVÁ, Petr KÖNIG a Hana ROSOLOVÁ. A multicentre study on unattended automated office blood pressure measurement in treated hypertensive patients. Blood Pressure. London: INFORMA HEALTHCARE, 2018, roč. 27, č. 4, s. 188-193. ISSN 0803-7051. Dostupné z: https://dx.doi.org/10.1080/08037051.2018.1425606.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název A multicentre study on unattended automated office blood pressure measurement in treated hypertensive patients
Autoři FILIPOVSKÝ, Jan (203 Česká republika), Jitka SEIDLEROVÁ (203 Česká republika, garant), Jiří CERAL (203 Česká republika), Petra VYSOČANOVÁ (203 Česká republika), Jiří ŠPÁC (203 Česká republika, domácí), Miroslav SOUČEK (203 Česká republika, domácí), Ivan ŘIHÁČEK (203 Česká republika, domácí), Markéta MATEŘÁNKOVÁ (203 Česká republika), Petr KÖNIG (203 Česká republika) a Hana ROSOLOVÁ (203 Česká republika).
Vydání Blood Pressure, London, INFORMA HEALTHCARE, 2018, 0803-7051.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 30201 Cardiac and Cardiovascular systems
Stát vydavatele Velká Británie a Severní Irsko
Utajení není předmětem státního či obchodního tajemství
Impakt faktor Impact factor: 2.292
Kód RIV RIV/00216224:14110/18:00103952
Organizační jednotka Lékařská fakulta
Doi http://dx.doi.org/10.1080/08037051.2018.1425606
UT WoS 000437328100002
Klíčová slova anglicky Automated blood pressure; ambulatory blood pressure monitoring; attended blood pressure; BPTru device; blood pressure measurement
Štítky 14110116, rivok
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnila: Soňa Böhmová, učo 232884. Změněno: 9. 2. 2019 19:48.
Anotace
Aims: Unattended automated office blood pressure (uAutoOBP) may eliminate white-coat effect. In the present study, we studied its relationships to attended office blood pressure (BP) and ambulatory BP monitoring (ABPM).Material and methods: Stable treated hypertensive subjects were examined in four Czech academic hypertension centres. uAutoOBP was measured with the BP Tru device; attended BP was measured six times: three times with auscultatory method (AuscOBP) by the physician followed optionally by three oscillometric measurements (OscOBP). ABPM was performed within one week from the clinical visit.Results: Data on 172 subjects aged 63.712.4 years with AuscOBP 127.6 +/- 12.1/77.6 +/- 10.0mm Hg are reported. uAutoOBP was by 8.5 +/- 9.0/3.0 +/- 6.1mm Hg lower than AuscOBP. The AuscOBP-uAutoOBP difference increased with the AuscOBP level and it did not depend on any other factor. OscOBP differed by 8.6 +/- 8.6/1.9 +/- 5.7mm Hg from uAutoOBP. 24-hour mean BP was by 4.2 +/- 12.1/3.5 +/- 7.8mm Hg lower than AuscOBP and by 4.3 +/- 11.0/0.5 +/- 6.9mm Hg higher than uAutoOBP; the correlation coefficients of 24-hour mean BP with AuscOBP and with uAutoOBP did not differ (p for difference.13). In the lowest BP group (systolic AuscOBP<120mm Hg or diastolic AuscOBP<70mm Hg), both AuscOBP and uAutoOBP were lower than 24-hour mean BP, while in the highest BP group (systolic AuscOBP140mm Hg or diastolic AuscOBP90mm Hg), they were higher.Conclusions: Compared to uAutoOBP, attended BP measurement gives higher values, both when measured with auscultatory or oscillometric method. Inter-individual variability of AutoOBP - uAuscOBP difference, as well of uAutoOBP - ABPM difference, is large. We did not prove that uAutoOBP would be associated to 24-hour ambulatory BP more closely than attended BP.