2018
Evaluation of FUS-2000 urine analyzer: analytical properties and particle recognition
BEŇOVSKÁ, Miroslava, Ondřej WIEWIORKA a Jana PINKAVOVAZákladní údaje
Originální název
Evaluation of FUS-2000 urine analyzer: analytical properties and particle recognition
Autoři
BEŇOVSKÁ, Miroslava (203 Česká republika, garant, domácí), Ondřej WIEWIORKA (203 Česká republika, domácí) a Jana PINKAVOVA (203 Česká republika)
Vydání
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, ABINGDON, INFORMA HEALTHCARE, 2018, 0036-5513
Další údaje
Jazyk
angličtina
Typ výsledku
Článek v odborném periodiku
Obor
30100 3.1 Basic medicine
Stát vydavatele
Velká Británie a Severní Irsko
Utajení
není předmětem státního či obchodního tajemství
Impakt faktor
Impact factor: 1.380
Kód RIV
RIV/00216224:14110/18:00103986
Organizační jednotka
Lékařská fakulta
UT WoS
000424952600023
Klíčová slova anglicky
Laboratory routines and automation; urinalysis; analytical performance; laboratory statistics; microscopy; medical biochemistry
Příznaky
Mezinárodní význam, Recenzováno
Změněno: 9. 2. 2019 22:55, Soňa Böhmová
Anotace
V originále
This study evaluates the performance of microscopic part of a hybrid analyzer FUS-2000 (Dirui Industrial Co., Changchun, China), its analytical properties and particle recognition. The evaluation of trueness, repeatability, detection limit, carry-over, linearity range and analytical stability was performed according to Dirui protocol guidelines designed by Dirui Company to guarantee the quality of the instrument. Trueness for low, medium and high-value concentrations was calculated with bias of 15.5, 4.7 and -6.6%, respectively. Detection limit of 5 Ery/l was confirmed. Coefficient of variation of 11.0, 5.2 and 3.8% was measured for within-run repeatability of low, medium and high concentration. Between-run repeatability for daily quality control had coefficient of variation of 3.0%. Carry-over did not exceed 0.05%. Linearity was confirmed for range of 0-16,000 particles/l (R-2=0.9997). The analytical stability had coefficient of variation of 4.3%. Out of 1258 analyzed urine samples, 362 positive were subjected to light microscopy urine sediment analysis and compared to the analyzer results. Cohen's kappa coefficients were calculated to express the concordance. Squared kappa coefficient was 0.927 (red blood cells), 0.888 (white blood cells), 0.908 (squamous epithelia), 0.634 (transitional epithelia), 0.628 (hyaline casts), 0.843 (granular casts) and 0.623 (bacteria). Single kappa coefficients were 0.885 (yeasts) and 0.756 (crystals), respectively. Aforementioned results show good analytical performance of the analyzer and tight agreement with light microscopy of urine sediment.