VONDRAK, Jiri, Dan MAREK, Jan VECERA, Klára BENEŠOVÁ a Petr VOJTISEK. Cardiac resynchronization therapy - A comparison of VV delay optimization by 3D echocardiography using systolic dyssynchrony index and QRS width assessment at 6 months after CRT implantation. cor et Vasa. AMSTERDAM: ELSEVIER SCIENCE BV, 2018, roč. 60, č. 4, s. "E367"-"E376", 10 s. ISSN 0010-8650. Dostupné z: https://dx.doi.org/10.1016/j.crvasa.2017.11.006.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Cardiac resynchronization therapy - A comparison of VV delay optimization by 3D echocardiography using systolic dyssynchrony index and QRS width assessment at 6 months after CRT implantation
Autoři VONDRAK, Jiri (203 Česká republika, garant), Dan MAREK (203 Česká republika), Jan VECERA (203 Česká republika), Klára BENEŠOVÁ (203 Česká republika, domácí) a Petr VOJTISEK (203 Česká republika).
Vydání cor et Vasa, AMSTERDAM, ELSEVIER SCIENCE BV, 2018, 0010-8650.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 30201 Cardiac and Cardiovascular systems
Stát vydavatele Nizozemské království
Utajení není předmětem státního či obchodního tajemství
Kód RIV RIV/00216224:14110/18:00104186
Organizační jednotka Lékařská fakulta
Doi http://dx.doi.org/10.1016/j.crvasa.2017.11.006
UT WoS 000442902500005
Klíčová slova anglicky 3D echocardiography; Dyssynchrony; Cardiac resynchronization therapy; Optimization; Systolic dyssynchrony index; Clinical outcome; Volumetric responders
Štítky 14119612, rivok
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnila: Soňa Böhmová, učo 232884. Změněno: 9. 2. 2019 21:15.
Anotace
Aim: The aim of this study was to compare the setting of interventricular (VV) delay by 3D echocardiography (3DE) using systolic dyssynchrony index (SDI) versus QRS width measurement in new cardiac resynchronization therapy (CRT) recipients. We observed the impact on the reduction of left ventricle volumes and increase in volumetric responders (defined as a = 15% reduction in left ventricular end-systolic volume (LVESv)) at the 6-month follow-up. Methods: We included 63 patients with recently implanted CRT in this open-label, randomized trial. Patients were randomized into two groups. VV delay was set by the QRS width in the group 1 (n 31) to obtain the narrowest QRS complex and by SDI in the group 2 (n 32) to achieve the lowest possible value. We evaluated LVESv, left ventricular ejection fraction (LVEF) by 3DE, before CRT implantation and at 6-month follow-up, in all patients. We also obtained clinical parameters and the level of NT-proBNP. Results: The second group showed only a trend toward greater reduction of LVESv (-33 +/- 55 ml vs. -48 +/- 43 ml; P 0.367), increase in LVEF (+ 7.3 +/- 10.9% vs. + 10.2 +/- 9.4%; P 0.210) and greater number of volumetric responders (14 vs. 18; P 0.612) compared with the group 1 at 6-month follow-up. There were also no significant differences in clinical outcomes and the level of NT-proBNP. Conclusion: Individual CRT optimization using SDI compared with QRS duration assessment did not reveal any significant differences in echocardiographic parameters and clinical outcomes at 6-month follow-up. (c) 2017 The Czech Society of Cardiology. Published by Elsevier Sp. z o. o. All rights reserved.
VytisknoutZobrazeno: 28. 7. 2024 00:09