GRUWEL, M.L.H., Peter LATTA, Anna WOJNA PELCZAR, S. WOLFSBERGER a Boguslaw TOMANEK. MR imaging of tissue near aneurysm clips using short- and zero time MR sequences. MEASUREMENT. OXFORD: ELSEVIER SCI LTD, 2018, roč. 130, DEC, s. 398-403. ISSN 0263-2241. Dostupné z: https://dx.doi.org/10.1016/j.measurement.2018.08.017.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název MR imaging of tissue near aneurysm clips using short- and zero time MR sequences
Autoři GRUWEL, M.L.H. (36 Austrálie), Peter LATTA (703 Slovensko, garant, domácí), Anna WOJNA PELCZAR (616 Polsko, domácí), S. WOLFSBERGER (40 Rakousko) a Boguslaw TOMANEK (616 Polsko, domácí).
Vydání MEASUREMENT, OXFORD, ELSEVIER SCI LTD, 2018, 0263-2241.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 20601 Medical engineering
Stát vydavatele Velká Británie a Severní Irsko
Utajení není předmětem státního či obchodního tajemství
Impakt faktor Impact factor: 2.791
Kód RIV RIV/00216224:14740/18:00101371
Organizační jednotka Středoevropský technologický institut
Doi http://dx.doi.org/10.1016/j.measurement.2018.08.017
UT WoS 000446464400037
Klíčová slova anglicky Magnetic resonance imaging; Susceptibility artifacts; Aneurysm clip
Štítky rivok
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnila: Mgr. Pavla Foltynová, Ph.D., učo 106624. Změněno: 13. 3. 2019 11:27.
Anotace
This study shows how 3D ultra-short (UTE) and zero echo time (ZTE) experiments could be used to minimize susceptibility artifacts caused by MR-compatible aneurysm clips. Aneurysm clips are used to stop or prevent an aneurysm from bleeding. Postoperative imaging is needed to detect incomplete clipping, de novo aneurysm formation or multiple aneurysms. MRI is a non-invasive, radiation-free, technique to diagnose aneurysms. Unfortunately, treatment assessment by MRI after surgical placement of an aneurysm clip is complicated due to the presence of the metal clip, resulting in susceptibility artifacts. The clip's high magnetic susceptibility causes severe, orientation dependent, variations in the local magnetic field. Often this results in pronounced MR image distortions, including signal voids and pile-ups. We show that ZTE and 3D-UTE experiments, using radial sampling, provide MR images with less distortions than sequences using slice selection. Due to the minimal delay after RF-excitation and a significant reduction in signal time evolution during acquisition, susceptibility artifacts can be reduced. (C) 2018 Elsevier Ltd. All rights reserved.
Návaznosti
GA15-12607S, projekt VaVNázev: Návrh a optimalizace pulzních sekvenci s ultrakrátkým echо-časem pro spolehlivou detekci obsahu myelinu v lidském mozku pomocí MR zobrazování.
Investor: Grantová agentura ČR, Návrh a optimalizace pulzních sekvenci s ultrakrátkým echо-časem pro spolehlivou detekci obsahu myelinu v lidském mozku pomocí MR zobrazování