KUČERA, Matěj, Jiří VOLÁNEK, Pavel KRÁSENSKÝ, Přemysl LUBAL, Zdeněk FARKA a Marta FARKOVÁ. Development of instrumentation for coulometric titrations. In Libuše Trnková. XIX. Workshop of Biophysical Chemists and Electrochemists. 1st edition. Brno: Masaryk University Press, 2019, s. 48-49. ISBN 978-80-210-9309-6.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Development of instrumentation for coulometric titrations
Autoři KUČERA, Matěj, Jiří VOLÁNEK, Pavel KRÁSENSKÝ, Přemysl LUBAL, Zdeněk FARKA a Marta FARKOVÁ.
Vydání 1st edition. Brno, XIX. Workshop of Biophysical Chemists and Electrochemists, od s. 48-49, 2 s. 2019.
Nakladatel Masaryk University Press
Další údaje
Originální jazyk angličtina
Typ výsledku Stať ve sborníku
Obor 10406 Analytical chemistry
Stát vydavatele Česká republika
Utajení není předmětem státního či obchodního tajemství
Forma vydání paměťový nosič (CD, DVD, flash disk)
WWW URL
Organizační jednotka Přírodovědecká fakulta
ISBN 978-80-210-9309-6
Klíčová slova anglicky electrochemistry; coulometry; titration; iodometry
Změnil Změnila: Mgr. Marie Šípková, DiS., učo 437722. Změněno: 25. 4. 2024 10:18.
Anotace
This contribution is focused on the development and testing of coulometric analyzer, which was employed for in situ generation of analytical agent. The goal of the project was construction of a new coulometer, which ensures the exact and precise automatic coulometric titration. Coulometry is one of the basic electroanalytical methods. Two main approaches are direct and indirect coulometry, based on monitoring of the current required to convert the substance of interest into a precisely defined product (direct), and/or the amount of current required to produce an equivalent amount of reagent which takes place in the reaction with selected analyte (indirect). The main advantage of indirect coulometry is the possibility of in situ generation of reactive unstable ions that may serve as titration agents. The newly constructed device is based on an integrated circuit with a connected screen and electrode connection inputs. The measurement set-up consists of a beaker in which the coulometric titration takes place, and a pair of generator Pt electrodes connected to a controllable DC source for coulometric analyzer. Reference and Pt indicator electrodes are utilized for potentiometric detection of equivalence point. The titrations in this work are based on oxidization of analyte by iodine, which was coulometrically generated from alkaline potassium iodide. Other additives in solution were oxalic acid, which ensures the stability of analyzed ascorbic acid, and starch, which serves as an indicator forming blue-colored inclusive compounds with iodine. Firstly, the sample solution of ascorbic acid was titrated and its oxidation over time was recorded. Oxidation of ascorbic acid occurs due to oxygen in the air, even in solution, where metal ions have the same effect. All titrations of ascorbic acid for different concentrations were evaluated by observing the blue color of starch-iodine inclusion complex. However, the color of titration solution is very weak at the equivalence point, and thus the end of titration is subjective. Secondly, the purity of sodium thiosulphate was checked using two ways of equivalence point estimation – (i) again using visual detection with starch, and (ii) by potentiometric indication using the second-derivative curve, which allowed to achieve significantly improved results under the optimized experimental conditions.
Návaznosti
LQ1601, projekt VaVNázev: CEITEC 2020 (Akronym: CEITEC2020)
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, CEITEC 2020
MUNI/A/1359/2018, interní kód MUNázev: Vývoj metod a instrumentace pro analýzu biologicky významných látek 2019
Investor: Masarykova univerzita, Vývoj metod a instrumentace pro analýzu biologicky významných látek 2019, DO R. 2020_Kategorie A - Specifický výzkum - Studentské výzkumné projekty
VytisknoutZobrazeno: 12. 7. 2024 13:45