ROHAN, Tomáš, Tomáš ANDRAŠINA, Peter MATKULČÍK, Tomáš JŮZA, Jan ZAVADIL, Iva SVOBODOVÁ, Dalibor ČERVINKA a Vlastimil VÁLEK. Simulation of the biliary metal stent recanalisation using endoluminal irreversible electroporation on porcine liver - in vivo experiment. In CIRSE 2019. 2019. ISSN 1432-086X. Dostupné z: https://dx.doi.org/10.1007/s00270-019-02282-x.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Simulation of the biliary metal stent recanalisation using endoluminal irreversible electroporation on porcine liver - in vivo experiment
Autoři ROHAN, Tomáš (203 Česká republika, garant, domácí), Tomáš ANDRAŠINA (703 Slovensko, domácí), Peter MATKULČÍK (703 Slovensko, domácí), Tomáš JŮZA (203 Česká republika, domácí), Jan ZAVADIL (203 Česká republika, domácí), Iva SVOBODOVÁ (203 Česká republika, domácí), Dalibor ČERVINKA (203 Česká republika) a Vlastimil VÁLEK (203 Česká republika, domácí).
Vydání CIRSE 2019, 2019.
Další údaje
Originální jazyk angličtina
Typ výsledku Konferenční abstrakt
Obor 30224 Radiology, nuclear medicine and medical imaging
Stát vydavatele Německo
Utajení není předmětem státního či obchodního tajemství
WWW URL
Impakt faktor Impact factor: 2.034
Kód RIV RIV/00216224:14110/19:00110835
Organizační jednotka Lékařská fakulta
ISSN 1432-086X
Doi http://dx.doi.org/10.1007/s00270-019-02282-x
Klíčová slova anglicky biliary intervention; stent recanalisation; irreversible electroporation
Štítky rivok
Příznaky Mezinárodní význam
Změnil Změnila: Mgr. Tereza Miškechová, učo 341652. Změněno: 21. 4. 2020 11:08.
Anotace
Purpose: To demonstrate feasibility of endoluminal irreversible electroporation using tubular catheter in biliary metal stent for recanalisation. Material and methods: 25 simulations of metal stent obstruction on porcine liver in vivo were performed by liver lower edge incision and its insertion in the biliary metal stent. In 14 cases endoluminal irreversible electroporation with tubular catheter prototype were performed, others were set as controls. Three diferent scenarios of IRE were simulated (one, two or zero electrodes in contact with metal stent). Ninety 100 us pulses with 300,650,1000 and 1300 Volts were used in each scenario. Electrical parameters, macroscopical changes, side efects were monitored during the procedure. Early complications were evaluated by CT immediately and 72 hours after the procedure, followed by samples harvesting. Results: There were no complications detected during the procedure and on following CTs, enduluminal IRE was succesful in all cases. Signifcant diference in impedance was observed between all scenario settings (p=0,02, Mann-Whitey). Short circuit occured immediately in 1300 V in all scenarios, in both cases of 1000 V with one electrode in contact with stent and in case where both electrodes were in contact with stent. The highest observed safe voltage for IRE in metal stent was 850 V/cm. Histopathological analysis showed signs of tissue necrosis which were not present in control samples. Conclusion: Endoluminal irreversible electroporation using tubular catheter seems safe and feasible method and has potential for non-thermal metal stent ingrowth treatment.
Návaznosti
MUNI/A/0996/2018, interní kód MUNázev: Infrared thermal imaging – neinvazivní zobrazovací metoda s využitím v klinickém prostředí (Akronym: IRT)
Investor: Masarykova univerzita, Infrared thermal imaging – neinvazivní zobrazovací metoda s využitím v klinickém prostředí, DO R. 2020_Kategorie A - Specifický výzkum - Studentské výzkumné projekty
MUNI/A/1574/2018, interní kód MUNázev: Onkologické radiologické intervence a jejich přínos v rámci komplexní onkologické léčby, srovnání s celorepublikovými výsledky onkologické léčby vybraných diagnóz II
Investor: Masarykova univerzita, Onkologické radiologické intervence a jejich přínos v rámci komplexní onkologické léčby, srovnání s celorepublikovými výsledky onkologické léčby vybraných diagnóz II, DO R. 2020_Kategorie A - Specifický výzkum - Studentské výzkumné projekty
VytisknoutZobrazeno: 8. 8. 2024 23:32