JŮZA, Tomáš, Tomáš ROHAN, Peter MATKULČÍK, Dalibor ČERVINKA, Veronika NOVOTNÁ, Vlastimil VÁLEK a Tomáš ANDRAŠINA. Thermal efect of irreversible electroporation a high-frequency irreversible electroporation inside metallic stent ex vivo. In CIRSE 2019. 2019. ISSN 1432-086X. Dostupné z: https://dx.doi.org/10.1007/s00270-019-02282-x.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Thermal efect of irreversible electroporation a high-frequency irreversible electroporation inside metallic stent ex vivo
Autoři JŮZA, Tomáš (203 Česká republika, garant, domácí), Tomáš ROHAN (203 Česká republika), Peter MATKULČÍK (703 Slovensko, domácí), Dalibor ČERVINKA (203 Česká republika), Veronika NOVOTNÁ (203 Česká republika), Vlastimil VÁLEK (203 Česká republika, domácí) a Tomáš ANDRAŠINA (703 Slovensko).
Vydání CIRSE 2019, 2019.
Další údaje
Originální jazyk angličtina
Typ výsledku Konferenční abstrakt
Obor 30224 Radiology, nuclear medicine and medical imaging
Stát vydavatele Německo
Utajení není předmětem státního či obchodního tajemství
WWW URL
Impakt faktor Impact factor: 2.034
Kód RIV RIV/00216224:14110/19:00110837
Organizační jednotka Lékařská fakulta
ISSN 1432-086X
Doi http://dx.doi.org/10.1007/s00270-019-02282-x
Klíčová slova anglicky irreversible electroporation; high-frequency irreversible electroporation; thermal effect; metal stent
Štítky rivok
Příznaky Mezinárodní význam
Změnil Změnila: Mgr. Tereza Miškechová, učo 341652. Změněno: 21. 4. 2020 11:16.
Anotace
Purpose: To compare electrical and thermal parameters of irreversible electroporation (IRE) and high-frequency irreversible electroporation (H-FIRE) using tubular electrode prototype in biliary metal stent. Material and methods: 3-electrode tubular catheter prototype was placed inside metallic biliary stent in ex vivo porcine liver model. 3 scenarios of stent occlusion using 2±1 mm thick piece of liver parenchyma were simulated. Only 2 electrodes were active in each time. IRE were performed by one hundred pulses at voltages 300V, 650V, 1000V and 1300V with 100 us pulse length. H-FIRE is a newer modifcation of electroporation methods using, in our case, 100 us burst of alternating polarity pulses of 2,5; 5; 10 and 25 us length, pause between pulses 0,5, 1 and 1,5 seconds with the same voltages as IRE. Values of electric current were measured. Thermal efects were monitored by infrared thermal imaging camera. Results: IRE and H-FIRE procedures were feasible in all settings with simulated obstruction of the voltages of 300 and 650V and in most of the cases with 1000 V. In higher voltages the safety limit of generator (12A) were exceeded. No signifcant diference in heating between IRE and H-FIRE was observed (p=0,49, Mann-Whitney). Maximal temperature increase in both methods was 7°C at 1000 V in model with one electrode in direct in contact with the stent and the other one with simulated obstruction. Conclusion: No signifcant diferences in thermal efects between IRE and H-FIRE delivering comparable amount of energy in metal stent were observed.
Návaznosti
MUNI/A/0996/2018, interní kód MUNázev: Infrared thermal imaging – neinvazivní zobrazovací metoda s využitím v klinickém prostředí (Akronym: IRT)
Investor: Masarykova univerzita, Infrared thermal imaging – neinvazivní zobrazovací metoda s využitím v klinickém prostředí, DO R. 2020_Kategorie A - Specifický výzkum - Studentské výzkumné projekty
MUNI/A/1574/2018, interní kód MUNázev: Onkologické radiologické intervence a jejich přínos v rámci komplexní onkologické léčby, srovnání s celorepublikovými výsledky onkologické léčby vybraných diagnóz II
Investor: Masarykova univerzita, Onkologické radiologické intervence a jejich přínos v rámci komplexní onkologické léčby, srovnání s celorepublikovými výsledky onkologické léčby vybraných diagnóz II, DO R. 2020_Kategorie A - Specifický výzkum - Studentské výzkumné projekty
VytisknoutZobrazeno: 8. 8. 2024 23:22