ASHRAFI, A.M., M. SYS, E. SEDLACKOVA, A.S. FARAG, V. ADAM, Jan PŘIBYL a L. RICHTERA. Application of the Enzymatic Electrochemical Biosensors for Monitoring Non-Competitive Inhibition of Enzyme Activity by Heavy Metals. Sensors. Basel: MDPI AG, 2019, roč. 19, č. 13, s. 2939-2952. ISSN 1424-8220. Dostupné z: https://dx.doi.org/10.3390/s19132939.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Application of the Enzymatic Electrochemical Biosensors for Monitoring Non-Competitive Inhibition of Enzyme Activity by Heavy Metals
Autoři ASHRAFI, A.M. (203 Česká republika), M. SYS (203 Česká republika), E. SEDLACKOVA (203 Česká republika), A.S. FARAG (203 Česká republika), V. ADAM (203 Česká republika), Jan PŘIBYL (203 Česká republika, garant, domácí) a L. RICHTERA (203 Česká republika).
Vydání Sensors, Basel, MDPI AG, 2019, 1424-8220.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 10406 Analytical chemistry
Stát vydavatele Švýcarsko
Utajení není předmětem státního či obchodního tajemství
WWW URL
Impakt faktor Impact factor: 3.275
Kód RIV RIV/00216224:14740/19:00112781
Organizační jednotka Středoevropský technologický institut
Doi http://dx.doi.org/10.3390/s19132939
UT WoS 000477045000100
Klíčová slova anglicky glucose oxidase; heavy metals; amperometric biosensor; non-competitive inhibition
Štítky CF NANO, rivok
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnila: Mgr. Pavla Foltynová, Ph.D., učo 106624. Změněno: 12. 5. 2020 13:29.
Anotace
The inhibition effect of the selected heavy metals (Ag+, Cd2+, Cu2+, and Hg2+) on glucose oxidase (GOx) enzyme from Aspergillus niger (EC 1.1.3.4.) was studied using a new amperometric biosensor with an electrochemical transducer based on a glassy carbon electrode (GCE) covered with a thin layer of multi-wall carbon nanotubes (MWCNTs) incorporated with ruthenium(IV) oxide as a redox mediator. Direct adsorption of multi-wall carbon nanotubes (MWCNTs) and subsequent covering with Nafion((R)) layer was used for immobilization of Gox. The analytical figures of merit of the developed glucose (Glc) biosensor are sufficient for determination of Glc in body fluids in clinical analysis. From all tested heavy metals, mercury(II) has the highest inhibition effect. However, it is necessary to remember that cadmium and silver ions also significantly inhibit the catalytic activity of Gox. Therefore, the development of Gox biosensors for selective indirect determination of each heavy metal still represents a challenge in the field of bioelectroanalysis. It can be concluded that amperometric biosensors, differing in the utilized enzyme, could find their application in the toxicity studies of various poisons.
Návaznosti
LM2015043, projekt VaVNázev: Česká infrastruktura pro integrativní strukturní biologii (Akronym: CIISB)
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, Czech Infrastructure for Integrative Structural Biology
LQ1601, projekt VaVNázev: CEITEC 2020 (Akronym: CEITEC2020)
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, CEITEC 2020
VytisknoutZobrazeno: 20. 8. 2024 21:26