POSPÍŠIL, David, Tomáš NOVOTNÝ, Jiří JARKOVSKÝ, Barbora FARKASOVÁ, Milan KOZÁK, Lubomír KŘIVAN, Jitka VLAŠÍNOVÁ, Petr KALA a Milan SEPŠI. Differences in right-to-left vs left-to-right interventricular conduction times in patients indicated to cardiac resynchronization therapy. PLoS One. San Francisco: Public Library Science, 2020, roč. 15, č. 2, s. 1-8. ISSN 1932-6203. Dostupné z: https://dx.doi.org/10.1371/journal.pone.0228731.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Differences in right-to-left vs left-to-right interventricular conduction times in patients indicated to cardiac resynchronization therapy
Autoři POSPÍŠIL, David (203 Česká republika, garant, domácí), Tomáš NOVOTNÝ (203 Česká republika, domácí), Jiří JARKOVSKÝ (203 Česká republika, domácí), Barbora FARKASOVÁ (203 Česká republika, domácí), Milan KOZÁK (203 Česká republika, domácí), Lubomír KŘIVAN (203 Česká republika, domácí), Jitka VLAŠÍNOVÁ (203 Česká republika, domácí), Petr KALA (203 Česká republika, domácí) a Milan SEPŠI (203 Česká republika, domácí).
Vydání PLoS One, San Francisco, Public Library Science, 2020, 1932-6203.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 30201 Cardiac and Cardiovascular systems
Stát vydavatele Spojené státy
Utajení není předmětem státního či obchodního tajemství
WWW URL
Impakt faktor Impact factor: 3.240
Kód RIV RIV/00216224:14110/20:00115393
Organizační jednotka Lékařská fakulta
Doi http://dx.doi.org/10.1371/journal.pone.0228731
UT WoS 000535220600026
Klíčová slova anglicky ELECTRICAL DELAY; HEART-FAILURE
Štítky 14110211, 14119612, rivok
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnila: Mgr. Tereza Miškechová, učo 341652. Změněno: 8. 2. 2021 15:11.
Anotace
Introduction Differences in conduction times from right ventricle to left ventricle and from left ventricle to right ventricle respectively were observed during biventricular devices implantation when changing pacing vector direction. In this article the phenomenon of interventricular conduction time differences is described and assessed in relationship to various clinical and electrophysiological parameters. Methods In 62 consecutive patients (9 females) interventricular conduction times between right and left ventricle in both directions were measured during cardiac resynchronization therapy device implantation procedure. Complex pacing protocol was performed. Results Investigated individuals was divided into 3 subgroups according to type of interventricular conduction pattern and statistically tested with various clinical data. Substantial differences in right-to-left vs left-to-right conduction times (> 5 ms, range 7-72 ms) were observed in 24 (39%) of all patients. They were more common in patients with dilated cardiomyopathy (20 of 38, 53%) compared to 4 (17%) of 24 patients with coronary artery disease (p = 0.011). The phenomenon occurred more often in hypertensive patients (p = 0.012). Other tested factors were nonsignificant. Conclusions There are almost no data on this topic. The occurrence of conduction difference phenomenon is quite common in dilated cardiomyopathy while it is rare in coronary artery disease. We assume the diffuse nature of the disease and the way of remodeling of myocardium play the main role. Knowledge of this phenomenon could be useful in personalized cardiac resynchronization therapy optimization.
Návaznosti
LM2018128, projekt VaVNázev: Český národní uzel Evropské sítě infrastruktur klinického výzkumu (Akronym: CZECRIN)