VEVERKOVÁ, Jana, Denisa BARTKOVA, Adam WEISER, Antonin DLOUHY, Petr BABULA, Petr ŠTĚPKA a Monika PÁVKOVÁ GOLDBERGOVÁ. Effect of Ni ion release on the cells in contact with NiTi alloys. Environmental Science and Pollution Research. Heidelberg: Springer Heidelberg, 2020, roč. 27, č. 8, s. 7934-7942. ISSN 0944-1344. Dostupné z: https://dx.doi.org/10.1007/s11356-019-07506-8.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Effect of Ni ion release on the cells in contact with NiTi alloys
Autoři VEVERKOVÁ, Jana (203 Česká republika, domácí), Denisa BARTKOVA (203 Česká republika), Adam WEISER (203 Česká republika), Antonin DLOUHY (203 Česká republika), Petr BABULA (203 Česká republika, domácí), Petr ŠTĚPKA (203 Česká republika, domácí) a Monika PÁVKOVÁ GOLDBERGOVÁ (203 Česká republika, garant, domácí).
Vydání Environmental Science and Pollution Research, Heidelberg, Springer Heidelberg, 2020, 0944-1344.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 10511 Environmental sciences
Stát vydavatele Německo
Utajení není předmětem státního či obchodního tajemství
WWW URL
Impakt faktor Impact factor: 4.223
Kód RIV RIV/00216224:14110/20:00114179
Organizační jednotka Lékařská fakulta
Doi http://dx.doi.org/10.1007/s11356-019-07506-8
UT WoS 000517084600014
Klíčová slova anglicky Nitinol; Cell; Biocompatibility; Biomaterial; Hydrogen modification
Štítky 14110515, 14110518, rivok
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnila: Mgr. Tereza Miškechová, učo 341652. Změněno: 4. 3. 2021 14:47.
Anotace
Nickel-titanium alloys have been used in medical applications for several years; however, biocompatibility of the material remains controversial. In the present study, the human umbilical vein endothelial cells (HUVEC) were cultured in contact with the nitinol used in two different heat treatment surface modifications-helium and hydrogen. The amount of Ni ions released from these alloys in contact with HUVEC was measured in media and in the cells by ICP-MS. An increased release of Ni ions was detected in He alloy compared with H-2 alloy modification with an elevation with the metal exposition duration (24 h vs. 72 h). The cells contained the Ni ions in both selected alloy modifications with the lower levels in H-2 alloys. To evaluate the potential of multiple metal applications, similar values were observed in media and in cell suspension for all surface modification combinations. The model analysis of effect of metal ion release on distant cells in the body showed that the concentration is interestingly similar to concentrations in cells in direct contact with the metal alloy. The cells are able to regulate the concentration of Ni ions within the cell. According to our best knowledge, the study for the first time describes the presence of Ni ions released from nitinol directly in the cells. In the case of the H-2 modification, the lowest levels of Ni ions were detected both in medium and in the cells, which likely increases the biocompatibility of the nitinol alloy.
Návaznosti
GA15-16336S, projekt VaVNázev: Intersticiální příměsi v tvarově-paměťových slitinách na bázi NiTi
Investor: Grantová agentura ČR, Intersticiální příměsi v tvarově-paměťových slitinách na bázi NiTi
MUNI/A/1453/2019, interní kód MUNázev: Příspěvek ke studiu molekulární, buněčné, tkáňové a systémové patofyziologie vybraných komplexních nemocí
Investor: Masarykova univerzita, Příspěvek ke studiu molekulární, buněčné, tkáňové a systémové patofyziologie vybraných komplexních nemocí, DO R. 2020_Kategorie A - Specifický výzkum - Studentské výzkumné projekty
VytisknoutZobrazeno: 18. 7. 2024 16:32