Assessment of left ventricular volumes and ejection fraction using ultra-low-dose thallium-201 SPECT on a CZT camera: a comparison with magnetic resonance imaging
SALA, Marek, Vladimír KINCL, Milan KAMÍNEK, Jiří VAŠINA, Jan MÁCHAL et. al.Základní údaje
Originální název
Assessment of left ventricular volumes and ejection fraction using ultra-low-dose thallium-201 SPECT on a CZT camera: a comparison with magnetic resonance imaging
Autoři
SALA, Marek (203 Česká republika, domácí), Vladimír KINCL (203 Česká republika, garant, domácí), Milan KAMÍNEK (203 Česká republika), Jiří VAŠINA (203 Česká republika), Jan MÁCHAL (203 Česká republika, domácí), Roman PANOVSKÝ (203 Česká republika, domácí), Věra FEITOVÁ (203 Česká republika), Lukáš OPATŘIL (203 Česká republika, domácí) a Tomas HOLECEK (203 Česká republika)
Vydání
Journal of Nuclear Cardiology, New York, Springer, 2022, 1071-3581
Další údaje
Jazyk
angličtina
Typ výsledku
Článek v odborném periodiku
Obor
30201 Cardiac and Cardiovascular systems
Stát vydavatele
Spojené státy
Utajení
není předmětem státního či obchodního tajemství
Odkazy
Impakt faktor
Impact factor: 2.400
Kód RIV
RIV/00216224:14110/22:00124847
Organizační jednotka
Lékařská fakulta
UT WoS
000532890600004
Klíčová slova anglicky
Left ventricular ejection fraction; Volumes; Myocardial mass; CZT; thallium
Příznaky
Mezinárodní význam, Recenzováno
Změněno: 20. 7. 2022 08:33, Mgr. Tereza Miškechová
Anotace
V originále
Background Cadmium-Zinc-Telluride (CZT) technology allows use of low activities of radiopharmaceuticals. The aim was to verify the values of left ventricular volume parameters, obtained via ultra-low-dose thallium Single Photon Emission Computed Tomography (SPECT) using a CZT camera. Methods and Results Forty-five patients referred for an assessment of myocardial perfusion or viability imaging were examined using CZT-SPECT and 1.5 T magnetic resonance (MRI) scanner. The ultra-low-dose protocol with 0.5 Mbq 201-Tl per kg of body weight was used. The values of end-systolic (ESV) and end-diastolic volumes (EDV), left ventricular ejection fraction (EF) and myocardial mass (MM) were assessed using both techniques. A very good correlation was found between the EF, ESV, and EDV values assessed with CZT-SPECT and cardiac magnetic resonance MRI; the Pearson coefficients were 0.86, 0.95, and 0.91, respectively. A moderate correlation was found for myocardial mass, r = 0.57. Compared to MRI, SPECT systematically overestimated ESV and MM, while it underestimates the EF, with P <= .001 in all cases. There was no difference in EDV estimation. Conclusions Left ventricular volumes and ejection fraction assessed via ultra-low-dose CZT-SPECT showed very good correlation with the values obtained by MRI. A moderate correlation was found for myocardial mass.