J 2021

On the limits of finite element models created from (micro)CT datasets and used in studies of bone-implant-related biomechanical problems

MARCIAN, Petr, Libor BORAK, Tomas ZIKMUND, Ladislava HORÁČKOVÁ, Jozef KAISER et. al.

Základní údaje

Originální název

On the limits of finite element models created from (micro)CT datasets and used in studies of bone-implant-related biomechanical problems

Autoři

MARCIAN, Petr (203 Česká republika), Libor BORAK (203 Česká republika, garant), Tomas ZIKMUND (203 Česká republika), Ladislava HORÁČKOVÁ (203 Česká republika, domácí), Jozef KAISER, Marek JOUKAL (203 Česká republika, domácí) a Jan WOLFF (276 Německo)

Vydání

Journal of the Mechanical Behavior of Biomedical Materials, Elsevier, Amsterdam, 2021, 1751-6161

Další údaje

Jazyk

angličtina

Typ výsledku

Článek v odborném periodiku

Obor

30404 Biomaterials

Stát vydavatele

Nizozemské království

Utajení

není předmětem státního či obchodního tajemství

Odkazy

Impakt faktor

Impact factor: 4.042

Kód RIV

RIV/00216224:14110/21:00121433

Organizační jednotka

Lékařská fakulta

UT WoS

000635269700003

Klíčová slova anglicky

Finite element method; Computational modeling; Computed tomography; Image resolution; Bone tissue; Mechanical strain

Štítky

Příznaky

Mezinárodní význam, Recenzováno
Změněno: 21. 4. 2021 09:40, Mgr. Tereza Miškechová

Anotace

V originále

Patient-specific approach is gaining a wide popularity in computational simulations of biomechanical systems. Simulations (most often based on the finite element method) are to date routinely created using data from imaging devices such as computed tomography which makes the models seemingly very complex and sophisticated. However, using a computed tomography in finite element calculations does not necessarily enhance the quality or even credibility of the models as these depend on the quality of the input images. Low-resolution (medical-)CT datasets do not always offer detailed representation of trabecular bone in FE models and thus might lead to incorrect calculation of mechanical response to external loading. The effect of image resolution on mechanical simulations of bone-implant interaction has not been thoroughly studied yet. In this study, the effect of image resolution on the modeling procedure and resulting mechanical strains in bone was analyzed on the example of cranial implant. For this purpose, several finite element models of bone interacting with fixationscrews were generated using seven computed tomography datasets of a bone specimen but with different image resolutions (ranging from micro-CT resolution of 25 ?m to medical-CT resolution of 1250 ?m). The comparative analysis revealed that FE models created from images of low resolution (obtained from medical computed tomography) can produce biased results. There are two main reasons: 1. Medical computed tomography images do not allow generating models with complex trabecular architecture which leads to substituting of the intertrabecular pores with a fictitious mass; 2. Image gray value distribution can be distorted resulting in incorrect mechanical properties of the bone and thus in unrealistic or even completely fictitious mechanical strains. The biased results of calculated mechanical strains can lead to incorrect conclusion, especially when bone-implant interaction is investigated. The image resolution was observed not to significantly affect stresses in the fixation screw itself; however, selection of bone material representation might result in significantly different stresses in the screw.

Návaznosti

LQ1601, projekt VaV
Název: CEITEC 2020 (Akronym: CEITEC2020)
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, CEITEC 2020
90110, velká výzkumná infrastruktura
Název: CzechNanoLab