J 2022

Two Bonebridge bone conduction hearing implant generations: audiological benefit and quality of hearing in children

ŠIKOLOVÁ, Soňa, Milan URÍK, Dagmar HOŠNOVÁ, Vít KRUNTORÁD, Michal BARTOŠ et. al.

Základní údaje

Originální název

Two Bonebridge bone conduction hearing implant generations: audiological benefit and quality of hearing in children

Autoři

ŠIKOLOVÁ, Soňa (203 Česká republika, domácí), Milan URÍK (703 Slovensko, garant, domácí), Dagmar HOŠNOVÁ (203 Česká republika, domácí), Vít KRUNTORÁD (203 Česká republika, domácí), Michal BARTOŠ (203 Česká republika, domácí), Oldřich MOTYKA (203 Česká republika) a Petr JABANDŽIEV (203 Česká republika, domácí)

Vydání

European Archives of Oto-Rhino-Laryngology, New York, Springer, 2022, 0937-4477

Další údaje

Jazyk

angličtina

Typ výsledku

Článek v odborném periodiku

Obor

30206 Otorhinolaryngology

Stát vydavatele

Spojené státy

Utajení

není předmětem státního či obchodního tajemství

Odkazy

Impakt faktor

Impact factor: 2.600

Kód RIV

RIV/00216224:14110/22:00124985

Organizační jednotka

Lékařská fakulta

UT WoS

000695739900001

Klíčová slova anglicky

Active transcutaneous bone conduction implant; Children; Quality of lif; Atresia; Localisation; Hearing outcomes

Štítky

Příznaky

Mezinárodní význam, Recenzováno
Změněno: 14. 2. 2024 14:15, Mgr. Tereza Miškechová

Anotace

V originále

Purpose The study aimed to evaluate audiological benefits, quality of hearing and safety of two Bonebridge generation: BCI601 and BCI602 (MED-EL, Innsbruck, Austria) in children. Methods Twelve children were implanted: five BCI601 and seven BCI602 comprising of ten conductive hearing loss, and two single sided deaf SSD subjects. Audiological outcomes tested were sound field audiometry, functional gain, speech recognition threshold (SRT50), speech recognition in noise (SPRINT) and localisation abilities. Subjective measures were Speech, Spatial and Qualities of Hearing Scale (SSQ12). Results The mean FG with the BCI601 was 25.0 dB and with the BCI602 28.0 dB. The benefit in SRT50 was 23.2 dB and 33.8 dB, respectively. The mean benefit in SPRINT was 15% and 6.7% and the localisation ability improved from 33.3° to 16° and from 26.2° to 17.6°, respectively. The two SSD subjects reported a FG of 17 dB, a benefit in SRT50 of 22.5 and a benefit in SPRINT of 20%. Subjective outcomes improved significantly and even exceeded the values of their age-and sex matched normal hearing peers. One revision was reported: a retroauricular emphysema above the implant occurred 12 months post-OP, it was resolved operatively with the implant still being functional. Conclusion The pediatric cohort reports significant audiological benefit, even exceeding that of the age- and sex matched control. The combination of the high safety and audiological benefit makes the Bonebridge a comfortable and effective option in hearing rehabilitation in children.

Návaznosti

MUNI/A/1118/2020, interní kód MU
Název: Ztráta sluchu v dětském věku - příčiny, diagnostika, možnosti rehabilitace V.
Investor: Masarykova univerzita, Ztráta sluchu v dětském věku - příčiny, diagnostika, možnosti rehabilitace V.
MUNI/A/1132/2021, interní kód MU
Název: Ztráta sluchu v dětském věku - příčiny, diagnostika, možnosti rehabilitace VI.