KANICKÝ, Viktor a Jean - Michel MERMET. Selection of internal standards for analysis of silicate rocks and limestones by laser ablation inductively coupled plasma atomic emission spectrometry. Fresenius Journal of Analytical Chemistry. Berlin: Springer-Verlag, 1996, roč. 355, 7/8, s. 887-888. ISSN 0937-0633.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Selection of internal standards for analysis of silicate rocks and limestones by laser ablation inductively coupled plasma atomic emission spectrometry
Autoři KANICKÝ, Viktor (203 Česká republika, garant) a Jean - Michel MERMET (250 Francie).
Vydání Fresenius Journal of Analytical Chemistry, Berlin, Springer-Verlag, 1996, 0937-0633.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 10406 Analytical chemistry
Stát vydavatele Německo
Utajení není předmětem státního či obchodního tajemství
Kód RIV RIV/00216224:14310/96:00000026
Organizační jednotka Přírodovědecká fakulta
Klíčová slova anglicky Inductively Coupled Plasma; Atomic Emission Spectrometry; Laser Ablation; Silicates; Lithiumtetraborate Fusion;
Štítky atomic emission spectrometry, inductively coupled plasma, laser ablation, Lithiumtetraborate fusion, silicates
Příznaky Recenzováno
Změnil Změnil: prof. RNDr. Viktor Kanický, DrSc., učo 408. Změněno: 29. 6. 2007 09:43.
Anotace
Included (Si, Ca) and added (Li, B) internal standards (IS) have been used comparatively in LA-ICP AES of silicates and limestones to improve both precision and accuracy. A Q-switched Nd:YAG laser (355 nm, 10 Hz, 10 mJ per shot) has been applied for ablation, Samples have been prepared by fusion with Li2B4O7 and measured using a Perkin-Elmer Optima 3000 ICP system, Both types of IS have given a relative standard deviation (RSD) less than 2-1%, which improved the repeatability by a factor of 2-10 and calibration graphs have been linear over the whole concentration range.
Anotace česky
Included (Si, Ca) and added (Li, B) internal standards (IS) have been used comparatively in LA-ICP AES of silicates and limestones to improve both precision and accuracy. A Q-switched Nd:YAG laser (355 nm, 10 Hz, 10 mJ per shot) has been applied for ablation, Samples have been prepared by fusion with Li2B4O7 and measured using a Perkin-Elmer Optima 3000 ICP system, Both types of IS have given a relative standard deviation (RSD) less than 2-1%, which improved the repeatability by a factor of 2-10 and calibration graphs have been linear over the whole concentration range.