Selection of internal standards for analysis of silicate rocks and limestones by laser ablation inductively coupled plasma atomic emission spectrometry

KANICKÝ, Viktor a Jean - Michel MERMET

Základní údaje

Originální název

Selection of internal standards for analysis of silicate rocks and limestones by laser ablation inductively coupled plasma atomic emission spectrometry

Autoři

KANICKÝ, Viktor (203 Česká republika, garant) a Jean - Michel MERMET (250 Francie)

Vydání

Fresenius Journal of Analytical Chemistry, Berlin, Springer-Verlag, 1996, 0937-0633

Další údaje

Jazyk

angličtina

Typ výsledku

Článek v odborném periodiku

Obor

10406 Analytical chemistry

Stát vydavatele

Německo

Utajení

není předmětem státního či obchodního tajemství

Kód RIV

RIV/00216224:14310/96:00000026

Organizační jednotka

Přírodovědecká fakulta

Klíčová slova anglicky

Inductively Coupled Plasma; Atomic Emission Spectrometry; Laser Ablation; Silicates; Lithiumtetraborate Fusion;

Příznaky

Recenzováno
Změněno: 29. 6. 2007 09:43, prof. RNDr. Viktor Kanický, DrSc.

Anotace

V originále

Included (Si, Ca) and added (Li, B) internal standards (IS) have been used comparatively in LA-ICP AES of silicates and limestones to improve both precision and accuracy. A Q-switched Nd:YAG laser (355 nm, 10 Hz, 10 mJ per shot) has been applied for ablation, Samples have been prepared by fusion with Li2B4O7 and measured using a Perkin-Elmer Optima 3000 ICP system, Both types of IS have given a relative standard deviation (RSD) less than 2-1%, which improved the repeatability by a factor of 2-10 and calibration graphs have been linear over the whole concentration range.

Česky

Included (Si, Ca) and added (Li, B) internal standards (IS) have been used comparatively in LA-ICP AES of silicates and limestones to improve both precision and accuracy. A Q-switched Nd:YAG laser (355 nm, 10 Hz, 10 mJ per shot) has been applied for ablation, Samples have been prepared by fusion with Li2B4O7 and measured using a Perkin-Elmer Optima 3000 ICP system, Both types of IS have given a relative standard deviation (RSD) less than 2-1%, which improved the repeatability by a factor of 2-10 and calibration graphs have been linear over the whole concentration range.