NOVOTNA, Veronika, R. CIPIN, Dalibor CERVINKA, Zdeněk STÁREK a Martin PEŠL. Numerical Modeling of Electroporation Process Using Endocardial Catheter. Online. In Jarm, T., Cvetkoska, A., Mahnič-Kalamiza, S., Miklavcic, D. 8th European Medical and Biological Engineering Conference. EMBEC 2020. IFMBE Proceedings. 80. vyd. Switzerland: Springer Nature Switzerland ag, 2021, s. 896-904. EMBEC. ISBN 978-3-030-64609-7. Dostupné z: https://dx.doi.org/10.1007/978-3-030-64610-3_100.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Numerical Modeling of Electroporation Process Using Endocardial Catheter
Autoři NOVOTNA, Veronika, R. CIPIN, Dalibor CERVINKA, Zdeněk STÁREK (203 Česká republika) a Martin PEŠL (203 Česká republika, domácí).
Vydání 80. vyd. Switzerland, 8th European Medical and Biological Engineering Conference. EMBEC 2020. IFMBE Proceedings, od s. 896-904, 9 s. EMBEC, 2021.
Nakladatel Springer Nature Switzerland ag
Další údaje
Originální jazyk angličtina
Typ výsledku Kapitola resp. kapitoly v odborné knize
Obor 30201 Cardiac and Cardiovascular systems
Stát vydavatele Švýcarsko
Utajení není předmětem státního či obchodního tajemství
Forma vydání elektronická verze "online"
WWW URL
Kód RIV RIV/00216224:14110/21:00128248
Organizační jednotka Lékařská fakulta
ISBN 978-3-030-64609-7
Doi http://dx.doi.org/10.1007/978-3-030-64610-3_100
Klíčová slova anglicky Electroporation Numerical modelling Endocardial catheter Heart
Změnil Změnila: Mgr. Tereza Miškechová, učo 341652. Změněno: 5. 4. 2023 08:29.
Anotace
This paper is about electroporation using an endocardial catheter. It is focused on numerical modeling of the electroporation process inspired by in vivo experiment. Two electroporation models are compared; model with electric conductivity dependent and independent on the electric field intensity. Further, it also demonstrates the effect of the flowing blood on obtained results. The electroporation effect in the homogeneous tissue is symmetrical unlike the effect in the inhomogeneous tissue. Blood flowing around the application electrode cools down the tissue but takes also majority of the current. The treatment has to be either longer or use higher voltage. The temperature was 40.7 °C for the model with electric conductivity dependent on the electric field intensity vs 45 °C with independent electric conductivity. Comparison of the simulated and measured data is not part of this paper.
Návaznosti
ROZV/23/LF9/2019, interní kód MUNázev: Mechanobiologické souvislosti účinku bronchodilatačních léčiv na vývoj získaných srdečních patologií - Analýza farmakologicky navozených změn kontrakce kardiomyocitů diferencovaných z kmenových buněk s pomocí mikroskopie atomárních sil
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, Mechanobiologické souvislosti účinku bronchodilatačních léčiv na vývoj získaných srdečních patologií - Analýza farmakologicky navozených změn kontrakce kardiomyocitů diferencovaných z kmenových buněk s pomocí mikroskopie atomárních sil, Interní rozvojové projekty