2005
Amperometric flow-injection determination of phenolic compounds using a biosensor with immobilised laccase, peroxidase and tyrosinase
SOLNÁ, Renáta a Petr SKLÁDALZákladní údaje
Originální název
Amperometric flow-injection determination of phenolic compounds using a biosensor with immobilised laccase, peroxidase and tyrosinase
Název česky
Amperometrický průtočný biosensor pro stanovení fenolických sloučenin s imobilisovanou lakasou, peroxidasou a tyrosinasou
Autoři
SOLNÁ, Renáta (203 Česká republika) a Petr SKLÁDAL (203 Česká republika, garant)
Vydání
Electroanalysis, Weinheim, Wiley-VCH Verlag GmbH, 2005, 1040-0397
Další údaje
Jazyk
angličtina
Typ výsledku
Článek v odborném periodiku
Obor
10600 1.6 Biological sciences
Stát vydavatele
Česká republika
Utajení
není předmětem státního či obchodního tajemství
Impakt faktor
Impact factor: 2.189
Kód RIV
RIV/00216224:14310/05:00014346
Organizační jednotka
Přírodovědecká fakulta
UT WoS
000234031500005
Klíčová slova anglicky
electrochemical biosensor; enzyme electrode; phenol; catechol; waste water
Změněno: 2. 1. 2006 13:27, prof. RNDr. Petr Skládal, CSc.
V originále
A screen-printed four-electrode sensor based on immobilization of laccase (Coriolus hirsutus), peroxidase (horseradish) and tyrosinase (mushroom) in the same array was developed for monitoring of phenols. The enzymes were immobilized onto a self-assembled monolayer (4-mercapto-1-butanol) modified gold surface via covalent attachment by epichlorohydrin coupling. The experimental conditions for simultaneous operation of the three enzymes were optimized based on catechol determination. The sensors were further applied for the amperometric detection of several substituted phenolic compounds, carried out using a single line flow-injection system. Hydrogen peroxide served as co-substrate for peroxidase. The limits of detection for phenols in aqueous solutions were in the micromolar range, one assay was completed in less than 5 min. The preliminary studies showed that the compatibility of the above mentioned enzyme array enabled the multielectrode biosensor to be applied to real samples including industrial wastewaters and surface waters.
Česky
Byl vyvinut multikanálový elektrochemický biosensor pro stanovení fenolických látek v odpadních vodách pomocí enzymů specifických k fenolům: lakasa, peroxidasy a tyrosinasa.
Návaznosti
MSM0021622413, záměr |
|