SKLÁDAL, Petr, Yveta SYMERSKÁ, Miroslav POHANKA, Bohuslav ŠAFÁŘ a Aleš MACELA. Electrochemical immunosensor for detection of Francisella tularensis. In Defense against bioterror. 1. vyd. Dordrecht, The Netherlands: Springer, 2005, s. 221-232. NATO Security through science ser. B. ISBN 1-4020-3385-0.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Electrochemical immunosensor for detection of Francisella tularensis
Název česky Elektrochemický imunosensor pro detekci Francisella tularensis
Autoři SKLÁDAL, Petr (203 Česká republika, garant), Yveta SYMERSKÁ (203 Česká republika), Miroslav POHANKA (203 Česká republika), Bohuslav ŠAFÁŘ (203 Česká republika) a Aleš MACELA (203 Česká republika).
Vydání 1. vyd. Dordrecht, The Netherlands, Defense against bioterror, od s. 221-232, 12 s. NATO Security through science ser. B, 2005.
Nakladatel Springer
Další údaje
Originální jazyk angličtina
Typ výsledku Kapitola resp. kapitoly v odborné knize
Obor 10600 1.6 Biological sciences
Stát vydavatele Nizozemské království
Utajení není předmětem státního či obchodního tajemství
Kód RIV RIV/00216224:14310/05:00015144
Organizační jednotka Přírodovědecká fakulta
ISBN 1-4020-3385-0
UT WoS 000236493200017
Klíčová slova anglicky Biosensor; Biodetection; Biological agent
Štítky Biodetection, Biological agent, biosensor
Změnil Změnil: prof. RNDr. Petr Skládal, CSc., učo 2202. Změněno: 28. 4. 2006 12:47.
Anotace
The immunosensor for rapid and sensitive detection of Francisella tularensis (Ft) was developed using screen-printed sensors with multiple (4) Au working electrodes. The sensing zones were modified with a biolayer of anti-Ft polyclonal antibody attached through the covalently immobilized Protein A. The sandwich assay format was employed with anti-Ft IgG-peroxidase conjugate as tracer. For electrochemical measurements, an automated portable detector was developed; it consisted of a digitally controlled potentiostat and four embedded miniperistaltic pumps. The flow-through measurement involved sequential incubations with sample (10 min), tracer (10 min) and substrate mixture (iodide and hydrogen peroxide). The enzymatically-produced iodine was detected electrochemically at - 50 mV. The limit of detection was 100 Ft cells/ml; the response was significantly higher (3x) when using the living cells compared to the heat inactivated ones.
Anotace česky
Byl vyvinut elektrochemický imunosensor pro rychlou detekci bakterie Francisella tularensis, patřící mezi biologické bojové prostředky.
Návaznosti
MSM0021622413, záměrNázev: Proteiny v metabolismu a při interakci organismů s prostředím
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, Proteiny v metabolismu a při interakci organismů s prostředím