STAIER, Florian, Heinz EIPEL, Petr MATULA, Alexei V EVSIKOV, Michal KOZUBEK, Christoph CREMER a Michael HAUSMANN. Micro axial tomography: A miniaturized, versatile stage device to overcome resolution anisotropy in fluorescence light microscopy. Rev. Sci. Instrum. USA: American Institute of Physica, 2011, roč. 82, č. 9, s. "093701-1"-"093701-8", 8 s. ISSN 0034-6748.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Micro axial tomography: A miniaturized, versatile stage device to overcome resolution anisotropy in fluorescence light microscopy
Autoři STAIER, Florian (276 Německo), Heinz EIPEL (276 Německo), Petr MATULA (203 Česká republika, domácí), Alexei V EVSIKOV (643 Rusko), Michal KOZUBEK (203 Česká republika, garant, domácí), Christoph CREMER (276 Německo) a Michael HAUSMANN (276 Německo).
Vydání Rev. Sci. Instrum. USA, American Institute of Physica, 2011, 0034-6748.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 20200 2.2 Electrical engineering, Electronic engineering, Information engineering
Stát vydavatele Spojené státy
Utajení není předmětem státního či obchodního tajemství
Impakt faktor Impact factor: 1.367
Kód RIV RIV/00216224:14330/11:00053149
Organizační jednotka Fakulta informatiky
UT WoS 000295621100023
Klíčová slova anglicky fluorescence; glass fibres; image reconstruction; image resolution; optical microscopes; optical microscopy; optical tomography; stepping motors
Štítky cbia-web
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnil: doc. RNDr. Petr Matula, Ph.D., učo 3019. Změněno: 12. 2. 2018 15:10.
Anotace
To overcome anisotropic optical resolution or point localization precision in 3D fluorescence microscopy, micro axial tomography was suggested which allows object tilting on the microscopic stage. Here, we present a miniaturized device which can be implemented in a motor driven microscope stage. The footprint of this device corresponds to a standard microscope slide. A special glass fiber can manually be adjusted in the object space of the microscope lens. A stepwise fiber rotation can be controlled by a miniaturized stepping motor incorporated into the device. Test particles were fixed onto glass fibers, optically localized with high precision, and automatically rotated to obtain views from different perspective angles under which distances of corresponding pairs of objects were determined. From these angle dependent distance values, the real 3D distance was calculated with a precision in the ten nanometer range using standard microscopic equipment. Only very few images registered under different rotation angles are sufficient for full 3D reconstruction.
Návaznosti
MSM0021622419, záměrNázev: Vysoce paralelní a distribuované výpočetní systémy
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, Vysoce paralelní a distribuované výpočetní systémy
2B06052, projekt VaVNázev: Vytipování markerů, screening a časná diagnostika nádorových onemocnění pomocí vysoce automatizovaného zpracování multidimenzionálních biomedicínských obrazů (Akronym: Biomarker)
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, Vytipování markerů, screening a časná diagnostika nádorových onemocnění pomocí vysoce automatizovaného zpracování multidimenzionálních biomedicínských obrazů
VytisknoutZobrazeno: 11. 5. 2024 07:52