J 2012

Electrochemical Immunosensor Coupled to Cyclone Air Sampler for Detection of Escherichia coli DH5 alpha in Bioaerosols

SKLÁDAL, Petr, Eva ŠVÁBENSKÁ, Jiří ŽERAVÍK, Jan PŘIBYL, Petra ŠIŠKOVÁ et. al.

Basic information

Original name

Electrochemical Immunosensor Coupled to Cyclone Air Sampler for Detection of Escherichia coli DH5 alpha in Bioaerosols

Authors

SKLÁDAL, Petr (203 Czech Republic, guarantor, belonging to the institution), Eva ŠVÁBENSKÁ (203 Czech Republic, belonging to the institution), Jiří ŽERAVÍK (203 Czech Republic, belonging to the institution), Jan PŘIBYL (203 Czech Republic, belonging to the institution), Petra ŠIŠKOVÁ (203 Czech Republic, belonging to the institution), Torbjorn TJARNHAGE (752 Sweden) and Inga GUSTAFSON (752 Sweden)

Edition

Electroanalysis, WILEY-VCH Verlag GmbH, 2012, 1040-0397

Other information

Language

English

Type of outcome

Článek v odborném periodiku

Field of Study

10600 1.6 Biological sciences

Country of publisher

United States of America

Confidentiality degree

není předmětem státního či obchodního tajemství

Impact factor

Impact factor: 2.817

RIV identification code

RIV/00216224:14310/12:00059725

Organization unit

Faculty of Science

UT WoS

000300931000011

Keywords (in Czech)

bioagen; sendvičové imunostanovení; biosensor; vzdálené měření; aerosolová komora

Keywords in English

Bioagents; Sandwich immunoassay; Biosensors; Remote sensing; Aerosol chamber

Tags

Tags

International impact, Reviewed
Změněno: 9/4/2013 21:11, Ing. Andrea Mikešková

Abstract

V originále

The portable detector ImmunoSMART was combined with the cyclone air sampler to detect the model bioagent E. coli DH5a in bioaerosols. The selected capture antibody was immobilized on gold electrodes, antibody-peroxidase conjugate served as a tracer. In buffer, the amperometric immunosensor measured 103 to 108 CFU/mL within 30 min. Detection in air was realized in a closed aerosol chamber. Automated dissemination of the cells, sampling and measurement allowed remote testing of the cyclone/immunodetector. The level of 150 CFU/L in air was clearly indicated (20 min analysis including 5 min sampling) and independently confirmed using the slit sampler and agar cultivation. The results seem promising for further extended tests in both aerosol chamber and field trial conditions.

In Czech

Přenosný imunosensor ImmunoSMART byl ve spojení s cyklonem použit pro imunochemické stanovení bioaerosolu z Escherichia coli DH5alfa v aerosolové komoře.

Links

MSM0021622413, plan (intention)
Name: Proteiny v metabolismu a při interakci organismů s prostředím
Investor: Ministry of Education, Youth and Sports of the CR, Proteins in metabolism and interaction of organisms with the environment