FILIPIK, Adam, Jiří JAN a Igor PETERLÍK. Time-of-Flight Based Calibration of an Ultrasonic Computed Tomography System. Radioengineering. PRAHA: SPOLECNOST PRO RADIOELEKTRONICKE INZENYRSTVI, 2012, roč. 21, č. 1, s. 533-544. ISSN 1210-2512.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Time-of-Flight Based Calibration of an Ultrasonic Computed Tomography System
Autoři FILIPIK, Adam (203 Česká republika), Jiří JAN (203 Česká republika) a Igor PETERLÍK (703 Slovensko, domácí).
Vydání Radioengineering, PRAHA, SPOLECNOST PRO RADIOELEKTRONICKE INZENYRSTVI, 2012, 1210-2512.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 20206 Computer hardware and architecture
Stát vydavatele Nizozemské království
Utajení není předmětem státního či obchodního tajemství
Impakt faktor Impact factor: 0.687
Kód RIV RIV/00216224:14330/12:00080255
Organizační jednotka Fakulta informatiky
UT WoS 000303135600023
Klíčová slova anglicky Ultrasonic computed tomography; ultrasonic transmission tomography; calibration of sensors; nonlinear optimisation; time-of-flight measurements
Změnil Změnil: RNDr. Pavel Šmerk, Ph.D., učo 3880. Změněno: 22. 5. 2015 06:45.
Anotace
The paper presents a novel method for calibration of measuring geometry and of individual signal delays of transducers in ultrasonic computed tomography (USCT) systems via computational processing of multiple time-of-flight measurements of ultrasonic (US) impulses. The positions and time-delay parameters of thousands of ultrasonic transducers inside the USCT tank are calibrated by this approach with a high precision required for the tomographic reconstruction; such accuracy cannot be provided by any other known method. Although utilising similar basic principles as the global positioning system (GPS), the method is importantly generalised in treating all transducer parameters as the to-be calibrated (floating) unknowns, without any a-priori known positions and delays. The calibration is formulated as a non-linear least-squares problem, minimizing the differences between the calculated and measured time-of-arrivals of ultrasonic pulses. The paper provides detailed derivation of the method, and compares two implemented approaches (earlier calibration of individual transducers with the new approach calibrating rigid transducer arrays) via detailed simulations, aimed at testing the convergence properties and noise robustness of both approaches. Calibration using real US signals is described and, as an illustration of the utility of the presented method, a comparison is shown of two image reconstructions using the tomographic US data from a concrete experimental USCT system measuring a 3D phantom, without and after the calibration.