2017
Determination of Zinc, Cadmium, Lead, Copper and Silver Using a Carbon Paste Electrode and a Screen Printed Electrode Modified with Chromium(III) Oxide
KOUDELKOVA, Zuzana; Tomáš SYROVY; Pavlína AMBROZOVA; Zdeněk MORAVEC; Lubomír KUBAC et al.Základní údaje
Originální název
Determination of Zinc, Cadmium, Lead, Copper and Silver Using a Carbon Paste Electrode and a Screen Printed Electrode Modified with Chromium(III) Oxide
Autoři
KOUDELKOVA, Zuzana; Tomáš SYROVY; Pavlína AMBROZOVA; Zdeněk MORAVEC; Lubomír KUBAC; David HYNEK; Lukáš RICHTERA a Vojtěch ADAM
Vydání
Sensors, Basel, MDPI AG, 2017, 1424-8220
Další údaje
Jazyk
angličtina
Typ výsledku
Článek v odborném periodiku
Obor
10406 Analytical chemistry
Stát vydavatele
Švýcarsko
Utajení
není předmětem státního či obchodního tajemství
Odkazy
Impakt faktor
Impact factor: 2.475
Označené pro přenos do RIV
Ano
Kód RIV
RIV/00216224:14310/17:00120390
Organizační jednotka
Přírodovědecká fakulta
UT WoS
EID Scopus
Klíčová slova anglicky
carbon paste; chromium; electrochemistry; heavy metals; screen-printed electrode; silver
Štítky
Příznaky
Mezinárodní význam, Recenzováno
Změněno: 9. 11. 2021 16:52, Mgr. Marie Novosadová Šípková, DiS.
Anotace
V originále
In this study, the preparation and electrochemical application of a chromium(III) oxide modified carbon paste electrode (Cr-CPE) and a screen printed electrode (SPE), made from the same material and optimized for the simple, cheap and sensitive simultaneous determination of zinc, cadmium, lead, copper and the detection of silver ions, is described. The limits of detection and quantification were 25 and 80 mu g center dot L-1 for Zn(II), 3 and 10 mu g center dot L-1 for Cd(II), 3 and 10 mu g center dot L-1 for Pb(II), 3 and 10 mu g center dot L-1 for Cu(II), and 3 and 10 mu g center dot L-1 for Ag(I), respectively. Furthermore, this promising modification was transferred to the screen-printed electrode. The limits of detection for the simultaneous determination of zinc, cadmium, copper and lead on the screen printed electrodes were found to be 350 mu g center dot L-1 for Zn(II), 25 mu g center dot L-1 for Cd(II), 3 mu g center dot L-1 for Pb(II) and 3 mu g center dot L-1 for Cu(II). Practical usability for the simultaneous detection of these heavy metal ions by the Cr-CPE was also demonstrated in the analyses of wastewaters.