SKOPALÍK, Josef, Jiří SEKORA, Martin PEŠL, Markéta BÉBAROVÁ, Olga ŠVECOVÁ, Tomáš PARÁK, Vratislav ČMIEL, Ivo PROVAZNÍK, Edita JEKLOVÁ a Josef MAŠEK. Upgrade of The Langendorff Apparatus Using The Infrared Thermo-Control System and An Intelligent Heater. Lékař a technika. Praha: Czech Society for Biomedical Engineering and Medical Informatics, roč. 50, č. 4, s. 137-141. ISSN 0301-5491. doi:10.14311/CTJ.2020.4.03. 2020.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Upgrade of The Langendorff Apparatus Using The Infrared Thermo-Control System and An Intelligent Heater
Autoři SKOPALÍK, Josef (203 Česká republika, garant), Jiří SEKORA (203 Česká republika), Martin PEŠL (203 Česká republika, domácí), Markéta BÉBAROVÁ (203 Česká republika, domácí), Olga ŠVECOVÁ (203 Česká republika, domácí), Tomáš PARÁK (203 Česká republika), Vratislav ČMIEL (203 Česká republika), Ivo PROVAZNÍK (203 Česká republika), Edita JEKLOVÁ (203 Česká republika) a Josef MAŠEK (203 Česká republika).
Vydání Lékař a technika, Praha, Czech Society for Biomedical Engineering and Medical Informatics, 2020, 0301-5491.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 30105 Physiology
Stát vydavatele Česká republika
Utajení není předmětem státního či obchodního tajemství
WWW URL
Kód RIV RIV/00216224:14110/20:00123174
Organizační jednotka Lékařská fakulta
Doi http://dx.doi.org/10.14311/CTJ.2020.4.03
Klíčová slova anglicky Langendorff Apparatus; pig heart; infrared; thermo-control; heater
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnila: doc. MUDr. Markéta Bébarová, Ph.D., učo 15000. Změněno: 8. 12. 2021 14:12.
Anotace
Biological experiments involving isolated organs and tissues demand precise temperature monitoring and regulation. An automatic temperature control system was proposed and optimised on real isolated swine hearts and the prototype is described in this work. The traditional Langendorff apparatus consists of a heart holder, a reservoir of perfusion solution flowing to aortic cannula and a heating bath allowing passive heat transfer to the reservoir of perfusion solution. The commercial infrared camera FLIR T62101 was added to this basic set-up and used for very precise monitoring of the temperature kinetic of the organ and connected with an electronic feedback loop, which allowed real-time and precise regulation of heat transfer from the heating bath to the perfusion solution and in turn indirectly to the heart tissue. This provides real time control and active regulation of the myocardial tissue temperature. The infrared camera was tested in several modes and several variants of detection were optimised for ideal measurement of the region of interest of the ex vivo organ. The kinetics of the temperature changes and temperature stability of the tissue were recorded and calibrated by external electronic thermometers (type Pt100, inserted in tissue). The time lapse from the hang-up of the hypo termed organ (30 °C) until optimal warming (37 °C) was less than eight minutes in the final instrument prototype. The final stability of the 37 °C tissue temperature was approved; the temperature fluctuation of left ventricle tissue was characterised as 36.8 ± 0.5 °C. This upgraded traditional instrument could be used in specific preclinical and clinical transplantation and analytical projects in future.