Percutaneous Endoluminal Radiofrequency Ablation of Occluded Biliary Metal Stent in Malignancy Using Monopolar Technique: A Feasibility Study
ROHAN, Tomáš, Tomáš ANDRAŠINA, Peter MATKULČÍK, Vladan BERNARD, Vlastimil VÁLEK et. al.Základní údaje
Originální název
Percutaneous Endoluminal Radiofrequency Ablation of Occluded Biliary Metal Stent in Malignancy Using Monopolar Technique: A Feasibility Study
Autoři
ROHAN, Tomáš (203 Česká republika, garant, domácí), Tomáš ANDRAŠINA (703 Slovensko, domácí), Peter MATKULČÍK (703 Slovensko, domácí), Vladan BERNARD (203 Česká republika, domácí) a Vlastimil VÁLEK (203 Česká republika, domácí)
Vydání
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, NEW YORK, SPRINGER, 2022, 0174-1551
Další údaje
Jazyk
angličtina
Typ výsledku
Článek v odborném periodiku
Obor
30224 Radiology, nuclear medicine and medical imaging
Stát vydavatele
Švýcarsko
Utajení
není předmětem státního či obchodního tajemství
Odkazy
Impakt faktor
Impact factor: 2.900
Kód RIV
RIV/00216224:14110/22:00125706
Organizační jednotka
Lékařská fakulta
UT WoS
000761841200001
Klíčová slova anglicky
Biliary malignancy; Biliary stenosis; Intraductal ablation; Metal stent; Monopolar technique; Radiofrequency ablation
Příznaky
Mezinárodní význam, Recenzováno
Změněno: 20. 1. 2023 10:46, Mgr. Tereza Miškechová
Anotace
V originále
Purpose: To prove feasibility and safety of percutaneous endoluminal radiofrequency ablation (eRFA) using a monopolar approach in treatment of occluded biliary stent in malignancy. Materials and Methods: The study included 11 patients with occluded biliary metal stent that had been implanted due to malignant biliary obstruction. All underwent metal stent recanalization by percutaneous eRFA in monopolar setting. Sixteen eRFA procedures were performed under fluoroscopic guidance with an EndoHPB 8F radiofrequency ablation catheter. The effect of stent recanalization was assessed based upon change from pre- to post-procedural diameter of the patent lumen of the metal stent (Wilcoxon test), primary and secondary stent patency (compared by log-rank test), catheter-free period, and overall survival. Adverse events were evaluated according to Common Terminology Criteria for Adverse Events (CTCEA) 4.0. Results: Recanalization of the metal stent by monopolar radiofrequency ablation was successful in all 11 patients. Diameter of the patent lumen of the stent significantly widened after the eRFA inside the stent (median 2 vs. 7 mm, p = 0.003). Grade 1 complications were observed in one-third of procedures. Median stent patency after recanalization by eRFA was non-inferior to primary metal stent patency (154 vs. 161 days, p = 0.27). Median catheter-free survival and overall survival after stent recanalization were 149 and 210 days, respectively. Conclusion: Endoluminal radiofrequency ablation in monopolar setting was shown to be a feasible and safe method for recanalization of occluded biliary metal stents.