Percutaneous Endoluminal Radiofrequency Ablation of Occluded Biliary Metal Stent in Malignancy Using Monopolar Technique: A Feasibility Study

ROHAN, Tomáš, Tomáš ANDRAŠINA, Peter MATKULČÍK, Vladan BERNARD, Vlastimil VÁLEK et. al.

Základní údaje

Originální název

Percutaneous Endoluminal Radiofrequency Ablation of Occluded Biliary Metal Stent in Malignancy Using Monopolar Technique: A Feasibility Study

Autoři

ROHAN, Tomáš (203 Česká republika, garant, domácí), Tomáš ANDRAŠINA (703 Slovensko, domácí), Peter MATKULČÍK (703 Slovensko, domácí), Vladan BERNARD (203 Česká republika, domácí) a Vlastimil VÁLEK (203 Česká republika, domácí)

Vydání

CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, NEW YORK, SPRINGER, 2022, 0174-1551

Další údaje

Jazyk

angličtina

Typ výsledku

Článek v odborném periodiku

Obor

30224 Radiology, nuclear medicine and medical imaging

Stát vydavatele

Švýcarsko

Utajení

není předmětem státního či obchodního tajemství

Odkazy

Impakt faktor

Impact factor: 2.900

Kód RIV

RIV/00216224:14110/22:00125706

Organizační jednotka

Lékařská fakulta

UT WoS

000761841200001

Klíčová slova anglicky

Biliary malignancy; Biliary stenosis; Intraductal ablation; Metal stent; Monopolar technique; Radiofrequency ablation

Štítky

Příznaky

Mezinárodní význam, Recenzováno
Změněno: 20. 1. 2023 10:46, Mgr. Tereza Miškechová

Anotace

V originále

Purpose: To prove feasibility and safety of percutaneous endoluminal radiofrequency ablation (eRFA) using a monopolar approach in treatment of occluded biliary stent in malignancy. Materials and Methods: The study included 11 patients with occluded biliary metal stent that had been implanted due to malignant biliary obstruction. All underwent metal stent recanalization by percutaneous eRFA in monopolar setting. Sixteen eRFA procedures were performed under fluoroscopic guidance with an EndoHPB 8F radiofrequency ablation catheter. The effect of stent recanalization was assessed based upon change from pre- to post-procedural diameter of the patent lumen of the metal stent (Wilcoxon test), primary and secondary stent patency (compared by log-rank test), catheter-free period, and overall survival. Adverse events were evaluated according to Common Terminology Criteria for Adverse Events (CTCEA) 4.0. Results: Recanalization of the metal stent by monopolar radiofrequency ablation was successful in all 11 patients. Diameter of the patent lumen of the stent significantly widened after the eRFA inside the stent (median 2 vs. 7 mm, p = 0.003). Grade 1 complications were observed in one-third of procedures. Median stent patency after recanalization by eRFA was non-inferior to primary metal stent patency (154 vs. 161 days, p = 0.27). Median catheter-free survival and overall survival after stent recanalization were 149 and 210 days, respectively. Conclusion: Endoluminal radiofrequency ablation in monopolar setting was shown to be a feasible and safe method for recanalization of occluded biliary metal stents.

Návaznosti

MUNI/A/1388/2021, interní kód MU
Název: Význam radiologických intervencí a pokročilých zobrazovacích metod v diagnostice a léčbě pacientů s maligním onemocněním
Investor: Masarykova univerzita, Význam radiologických intervencí a pokročilých zobrazovacích metod v diagnostice a léčbě pacientů s maligním onemocněním
NU21-08-00561, projekt VaV
Název: Využití ireverzibilní elektroporace a vysokofrekvenční ireverzibilní elektroporace k léčbě okluzí metalických stentů
Investor: Ministerstvo zdravotnictví ČR, Využití ireverzibilní elektroporace a vysokofrekvenční ireverzibilní elektroporace k léčbě okluzí metalických stentů, Podprogram 1 - standardní