KANICKÝ, Viktor, Vítězslav OTRUBA a Jean-Michel MERMET. Comparison of some analytical performance characteristic in inductively coupled plasma spectrometry of platinum group metals and gold. Online. Talanta. Amsterdam: Elsevier Science, 1999, roč. 1999, č. 4, s. 859-866, 7 s. ISSN 0039-9140. [citováno 2024-04-24]
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Comparison of some analytical performance characteristic in inductively coupled plasma spectrometry of platinum group metals and gold
Autoři KANICKÝ, Viktor (203 Česká republika, garant), Vítězslav OTRUBA (203 Česká republika) a Jean-Michel MERMET (250 Francie)
Vydání Talanta, Amsterdam, Elsevier Science, 1999, 0039-9140.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 10406 Analytical chemistry
Stát vydavatele Nizozemské království
Utajení není předmětem státního či obchodního tajemství
Impakt faktor Impact factor: 1.185
Kód RIV RIV/00216224:14310/99:00000888
Organizační jednotka Přírodovědecká fakulta
Klíčová slova anglicky noble metals; platinum; gold; inductively coupled plasma spectrometry
Štítky gold, inductively coupled plasma spectrometry, noble metals, platinum
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnil: prof. RNDr. Viktor Kanický, DrSc., učo 408. Změněno: 29. 6. 2007 10:13.
Anotace
The limits of detection, precision and matrix effects in the inductively coupled plasma spectrometry of platinum group metals (PGMs) and gold were measured and evaluated for four ICP-AES and one ICP-MS instrument. The sample matrix was a cationic surfactant used for the PGMs and gold preconcentration on a modified silica gel (C18). A sorption of ion associates of PGMs and gold chlorocomplexes with the cation of onium salt N(1-carbaethoxypentadecyl)-trimethyl ammonium bromide was considered. The calibration curves, limits of detection and matrix effects were evaluated in the presence of 0.003 mol dm(-3) of onium salt (1.3 mg cm(-3)) and 0.1 mol dm(-3) HCl. The values of limits of detection (3 sigma(bl)) of PGMs for all axial ICP instruments were mostly below 10 ng cm(-3). Lateral observation on dual view ICP instrument yielded only 3 times higher detection limits in comparison to the axial mode of the same spectrometer and the detection limits for ICP-MS instrument were on the levels of units or tens of pg cm(-3). These limits of detection did not significantly differ from values obtained with pure solutions. Matrix effects in the presence of onium salt did not exceed 12% depression in the analytical signals. Besides the coefficients of correlation, the uncertainties on centroids of concentrations were calculated for calibration graphs obtained by linear regression. (C) 1999 Elsevier Science B.V. All rights reserved.
Anotace česky
The limits of detection, precision and matrix effects in the inductively coupled plasma spectrometry of platinum group metals (PGMs) and gold were measured and evaluated for four ICP-AES and one ICP-MS instrument. The sample matrix was a cationic surfactant used for the PGMs and gold preconcentration on a modified silica gel (C18). A sorption of ion associates of PGMs and gold chlorocomplexes with the cation of onium salt N(1-carbaethoxypentadecyl)-trimethyl ammonium bromide was considered. The calibration curves, limits of detection and matrix effects were evaluated in the presence of 0.003 mol dm(-3) of onium salt (1.3 mg cm(-3)) and 0.1 mol dm(-3) HCl. The values of limits of detection (3 sigma(bl)) of PGMs for all axial ICP instruments were mostly below 10 ng cm(-3). Lateral observation on dual view ICP instrument yielded only 3 times higher detection limits in comparison to the axial mode of the same spectrometer and the detection limits for ICP-MS instrument were on the levels of units or tens of pg cm(-3). These limits of detection did not significantly differ from values obtained with pure solutions. Matrix effects in the presence of onium salt did not exceed 12% depression in the analytical signals. Besides the coefficients of correlation, the uncertainties on centroids of concentrations were calculated for calibration graphs obtained by linear regression. (C) 1999 Elsevier Science B.V. All rights reserved.
Návaznosti
MSM 143100003, záměrNázev: Studium plazmochemických reakcí v neizotermickém nízkoteplotním plazmatu a jeho interakcí s povrchem pevných látek
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, Studium plazmochemických reakcí v neizotermickém nízkoteplotním plazmatu a jeho interakcí s povrchem pevných látek
VS97020, projekt VaVNázev: Laboratoř plazmových zdrojů pro chemickou analýzu
Investor: Ministerstvo školství, mládeže a tělovýchovy ČR, Laboratoř plazmových zdrojů pro chemickou analýzu
VytisknoutZobrazeno: 24. 4. 2024 03:22