MATKULČÍK, Peter, Martin HEMZAL, Dalibor ČERVINKA, Veronika NOVOTNÁ, Tomáš ROHAN a Tomáš ANDRAŠINA. Endoluminal IRE for recanalisation of an occluded metal stent - mathematical model. In CIRSE 2022. 2022. Dostupné z: https://dx.doi.org/10.1007/s00270-022-03246-4.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Endoluminal IRE for recanalisation of an occluded metal stent - mathematical model
Autoři MATKULČÍK, Peter (703 Slovensko, garant, domácí), Martin HEMZAL (203 Česká republika), Dalibor ČERVINKA (203 Česká republika), Veronika NOVOTNÁ (203 Česká republika), Tomáš ROHAN (203 Česká republika, domácí) a Tomáš ANDRAŠINA (703 Slovensko, domácí).
Vydání CIRSE 2022, 2022.
Další údaje
Originální jazyk angličtina
Typ výsledku Konferenční abstrakt
Obor 30224 Radiology, nuclear medicine and medical imaging
Stát vydavatele Španělsko
Utajení není předmětem státního či obchodního tajemství
WWW URL
Kód RIV RIV/00216224:14110/22:00129689
Organizační jednotka Lékařská fakulta
Doi http://dx.doi.org/10.1007/s00270-022-03246-4
Klíčová slova anglicky irreversible electroporation; metal stent; recanalisation; stent occlusion; mathematical model
Příznaky Mezinárodní význam
Změnil Změnil: MUDr. Tomáš Rohan, Ph.D., učo 358905. Změněno: 27. 9. 2022 13:21.
Anotace
Purpose: To analyze heat distribution and ablation zones formed by endoluminal irreversible electroporation inside an occluded metallic stent based on 3D mathematical tissue model simulations. Materials and methods: The software utilizing finite element method (COMSOL Multiphysics) for 3D simulations, 2D and 3D calculations were used to define the extent of irreversible electroporation and thermal ablation zones in different IRE protocols (100 100µs pulses, 1Hz, 300V,650V,1000V,1300V). Endoluminal IRE procedure for recanalization of metal stent using the 3-electrode IRE catheter was simulated in mathematical model of perfused liver tissue. Multiple stents with specific mesh design by different manufactures (uncovered stents, Boston Scientific, Ella-CS, Micro-Tech medical, S&G Biotech) were analyzed. The obstruction of metal stent was simulated by 1,5mm layer of interposed tissue. Results were visualized as the color-coded distribution of electrical potential, electric field intensity and thermal stress in the tissue model. Results: There was no statistically significant difference between various stent mesh designs regarding electric current in interposed tissue and increase of temperature in the tissue surrounding the stent (p>0.6; Mann-Whitney test). Mean volume of IRE zone using 1300V protocol was 10.9-11.7% larger (184-189 vs 166-169mm^3) and the maximal temperature increase was 67.7-68.8% higher (15.2-15.6 vs 9.0-9.3°C) compared to IRE using 1000V. Conclusion: The different mesh design of biliary metal stent did not lead to significant difference of electric current and temperature increase, so the model seems to be universally usable across all analyzed uncovered metal stents.
Návaznosti
MUNI/A/1388/2021, interní kód MUNázev: Význam radiologických intervencí a pokročilých zobrazovacích metod v diagnostice a léčbě pacientů s maligním onemocněním
Investor: Masarykova univerzita, Význam radiologických intervencí a pokročilých zobrazovacích metod v diagnostice a léčbě pacientů s maligním onemocněním
NU21-08-00561, projekt VaVNázev: Využití ireverzibilní elektroporace a vysokofrekvenční ireverzibilní elektroporace k léčbě okluzí metalických stentů
Investor: Ministerstvo zdravotnictví ČR, Využití ireverzibilní elektroporace a vysokofrekvenční ireverzibilní elektroporace k léčbě okluzí metalických stentů, Podprogram 1 - standardní
VytisknoutZobrazeno: 27. 7. 2024 14:31