HORSKÁ, Jana, Aleš HRDLIČKA, Jitka HEGROVÁ, Martina BUCKOVÁ, David PROCHAZKA, Pavel POŘÍZKA, Viktor KANICKÝ a Jozef KAISER. Study of spectral lines behaviour with a double pulse 1064–1064 nm laser-induced breakdown spectroscopy system at the direct analysis of algae on the filter. Spectrochimica Acta Part B: Atomic Spectroscopy. Amsterdam: Elsevier, 2024, roč. 213, March 2024, s. 106883-106890. ISSN 0584-8547. Dostupné z: https://dx.doi.org/10.1016/j.sab.2024.106883.
Další formáty:   BibTeX LaTeX RIS
Základní údaje
Originální název Study of spectral lines behaviour with a double pulse 1064–1064 nm laser-induced breakdown spectroscopy system at the direct analysis of algae on the filter
Autoři HORSKÁ, Jana (203 Česká republika, domácí), Aleš HRDLIČKA (203 Česká republika, domácí), Jitka HEGROVÁ (203 Česká republika), Martina BUCKOVÁ (203 Česká republika), David PROCHAZKA (203 Česká republika), Pavel POŘÍZKA (203 Česká republika), Viktor KANICKÝ (203 Česká republika) a Jozef KAISER (203 Česká republika).
Vydání Spectrochimica Acta Part B: Atomic Spectroscopy, Amsterdam, Elsevier, 2024, 0584-8547.
Další údaje
Originální jazyk angličtina
Typ výsledku Článek v odborném periodiku
Obor 10406 Analytical chemistry
Stát vydavatele Nizozemské království
Utajení není předmětem státního či obchodního tajemství
WWW URL
Impakt faktor Impact factor: 3.300 v roce 2022
Organizační jednotka Přírodovědecká fakulta
Doi http://dx.doi.org/10.1016/j.sab.2024.106883
Klíčová slova česky spektroskopie laserem buzeného plazmatu (LIBS); filtr, řasy; optimalizace; diagnostika plazmatu
Klíčová slova anglicky Laser-induced breakdown spectroscopy; Algae; Filter; Optimization; Plasma diagnostics
Příznaky Mezinárodní význam, Recenzováno
Změnil Změnil: Mgr. Aleš Hrdlička, Ph.D., učo 22863. Změněno: 8. 4. 2024 12:26.
Anotace
The nickel and zinc contamination of algae on a cellulose filter was studied with double pulse 1064–1064 nm laser-induced breakdown spectroscopy (LIBS). Samples of the green alga Desmodesmus subspicatus were contaminated with zinc and nickel. The surface concentrations on the filter calculated from the reference Inductively Coupled Plasma Mass Spectrometry analysis were 25 and 64 ng cm−2 for nickel and zinc respectively. The algae are resting on the filter after drying and a double adhesive tape was used to fix the sample on the microscopic glass. During optimization method a compromise of 23–30 mJ pulse energies were set for maintaining a detectable or highest signal intensity of the analyte and not destroying the filter. Best conditions for maximum analyte signal and minimum deviation between the spectral behaviour of the analyte and the compared line of Na I, K I, Mg I, Mg II, Mn I, Mn II, Ca I, Ca II, and C I was 700 ns gate delay and 0 s interpulse delay. The deviation was rather more influenced with the signal-to-noise ratio of the analyte line than the closeness of the upper-level energies of the compared lines. The filter without algae showed systematically lower electron number density and total emissivity by units of percents than the filter with algae.
Návaznosti
MUNI/A/1298/2022, interní kód MUNázev: Základní a aplikovaný výzkum a vývoj metod chemické a fyzikálně chemické analýzy pro studium přírody a pokročilé technologie
Investor: Masarykova univerzita, Základní a aplikovaný výzkum a vývoj metod chemické a fyzikálně chemické analýzy pro studium přírody a pokročilé technologie
VytisknoutZobrazeno: 24. 5. 2024 05:00